background image

2

KEYS

The Table 1.2 lists the four front panel keys and the associated function. 

Table 1.2

Symbol

Key

Function

Press to enter or exit set-up mode. 

DOWN

UP

ENTER

Press to store the set parameter value and to scroll to the next 
parameter on the PAGE.  

Press  to  increase  the  parameter  value.  Pressing  once 
increases the value by one count; holding pressed speeds up 
the change. 

Press  to  decrease  the  parameter  value.  Pressing  once 
decreases the value by one count; keeping pressed speeds up 
the change. 

PAGE

 User Manual

Clave   

Table 1.1

Status

Indicator

HTR

AIR

Indicates Heater On/Off status.

Indicates Air Outlet valve On/Off status.

  ALM

Flashes while the Alarm is active.

  S

P

These two indicators together indicate the parameter name whose value 
is currently being displayed on the Lower Readout.

             (Start) message.

(The controller is waiting for a ‘Start’ command to 
initiate a new autoclave cycle.) 

    

  S

  P

  What Lower Readout Indicates

  OFF

  OFF

  ON

  ON

  ON

  OFF

  OFF

  ON

Control Setpoint.

Saturated steam pressure in Kg/cm²

 

or PSI.

Soak Time Counter.

Содержание claveX 48 x 48

Страница 1: ...User Manual Autoclave Controller Clave...

Страница 2: ...ELLAYOUT 1 2 BASIC OPERATION 3 3 SET UPMODE ACCESS OPERATION 6 4 INPUT OUTPUT CONFIGURATION PARAMETERS 8 5 SOAK TIMER PARAMETERS OPERATION DETAILS 9 6 SUPERVISORYPARAMETERS 11 7 ELECTRICALCONNECTIONS...

Страница 3: ...or the balance Soak Time value depending on the autoclave cycle state In Program Mode the Lower Readout displays parameter names prompts INDICATORS TheTable 1 1 lists each front panel LED and the asso...

Страница 4: ...Pressing once decreases the value by one count keeping pressed speeds up the change PAGE User Manual Clave Table 1 1 Status Indicator HTR AIR Indicates Heater On Off status Indicates Air Outlet valve...

Страница 5: ...icating that the SP value is being shown on the Lower Readout Both the heater and air outlet are maintained ON As soon as the Temperature reaches theAir Outlet SP the air outlet is turned OFF Step 2 T...

Страница 6: ...e deviates above the control setpoint by more than the set band value see Operator Page and Parameters 3 PV Error The Alarm activates if the measured temperature value crosses the specified Max or Min...

Страница 7: ...e AIR OUTLET SETPOINT The setpoint value at and above which the Air Outlet valve is kept closed to build saturated steam pressure The Setpoint value at which the autoclave temperature value is maintai...

Страница 8: ...s to the MAIN Display Mode 4 Press and release the ENTER key until the prompt for the required parameter appears on the Lower Readout The last parameter in the list rolls back to the first parameter 5...

Страница 9: ...h the Lock enabled The Figure 3 2 below illustrates the Locking procedure Figure 3 2 UnLocking Repeat the Locking procedure twice for unlocking MASTER LOCKING The controller facilitates locking all th...

Страница 10: ...Control and Air Outlet setpoints HYSTERESIS Sets a differential dead band between the ON and OFF heater states Keep it large enough to avoid frequent switching of the heater without losing the desire...

Страница 11: ...1 to 5 0 Default 1 0 Re Start Abort Continue Default Continue Refer end of this section for detailed Soak Timer Operation Table 5 1 TIMER START BAND The temperature band around the control setpoint T...

Страница 12: ...on The timer is also provided with a Hold Band that can be enabled to make sure that the timer counts down only while the Temperature Value is within the Hold Band That is the timer pauses stops count...

Страница 13: ...esired rapid changes in the measured The higher the filter Temperature Value value the better the averaging but the slower the response to actual changes OFFSET FOR TEMPERATURE VALUE This parameter ad...

Страница 14: ...oth reading and writing No The Read Write parameters can only be accessed for reading That is the parameter values cannot be altered through serial communication Available for Utility Option Serial Co...

Страница 15: ...Use low resistance copper conductor leads of the same gauge and length Avoid joints in the cable Section 7 User Manual Clave 48 X 48 96 X 96 OUTPUT 1 Heater Control Output The Output 1 is factory conf...

Страница 16: ...ual Clave Line Neutral 12 L 13 N 2 Pole Isolating Switch Fuse Power Supply Terminal POWER SUPPLY Figure 7 6 Figure 7 4 a 18 17 16 NO C NC 16 17 18 48X48 96X96 As standard the controller accepts single...

Страница 17: ...a n e t Process Precision Instruments 101 Diamond Industrial Estate Navghar Vasai Road E Dist Palghar 401 210 Maharashtra India 0250 2391722 33 37 42 07498799226 09321985369 support ppiindia net sales...

Отзывы: