I / O CONFIGURATION PARAMETERS
Select the units for saturated steam pressure from Kg/cm and
²
PSI.
UNITS FOR PRESSURE
SETPOINT HIGH LIMIT
Use this limit to prevent accidental over settings of the ‘Control’
and ‘Air Outlet’ setpoints.
SETPOINT LOW LIMIT
Use this limit to prevent accidental under settings of the ‘Control’
and ‘Air Outlet’ setpoints.
HYSTERESIS
Sets a differential (dead) band between the ON and OFF heater
states. Keep it large enough to avoid frequent switching of the
heater without losing the desired control accuracy.
1 to 999°C or
0.1 to 99.9°C
(Default :
)
0.2
Min. Range for the
selected Input Type to
Setpoint High
(Default : 9
)
0.0
Setpoint Low to
Max. Range for the
selected Input Type
(Default :
)
150.0
Table 4.1
Parameter Description
Settings
(Default Value)
PSI
Kg/cm²
(Default :
²)
Kg/cm
Section 4
User Manual
Clave