The various parameters are arranged in different groups, called PAGES, depending upon the functions they represent. Each 

group is assigned a unique numeric value, called PAGE NUMBER, for its access.

The parameters are always presented in a fixed format: The Lower Readout displays the parameter prompt (Identification 

Name)  and  the  Upper  Readout  displays  the  set  value. The  parameters  appear  in  the  same  sequence  as  listed  in  their 

respective sections.

SET-UP MODE

The Set-up Mode allows the user to view and modify the parameter values. Follow the steps below for setting the parameter 

values: 

1.     Press and release PAGE key. The Lower Readout shows PAGE and the Upper Readout shows page number 0. Refer 

Figure 3.1.

2.     Use UP / DOWN keys to set the desired PAGE NUMBER. 

3.     Press and release ENTER key. The Lower Readout shows the prompt for the first parameter listed in the set PAGE and 

the Upper Readout shows its current value. If the entered PAGE NUMBER is invalid (contains no parameter list or any 

associated function), the controller reverts to the MAIN Display Mode.

4.     Press and release the ENTER key until the prompt for the required parameter appears on the Lower Readout. (The last 

parameter in the list rolls back to the first parameter). 

5.      Use  UP  /  DOWN  keys  to  adjust  the  parameter  value.  (The  display  flashes  if  UP  key  is  pressed  after  reaching  the  

maximum value or DOWN key is pressed after reaching the minimum value).

6.     Press and release the ENTER key. The new value gets stored in the controller’s non-volatile memory and the next 

parameter in the list is displayed. 

The Figure 3.1 illustrates the example of altering the value for the parameter  ‘Input Type’. 

Figure 3.1

Use UP/DOWN 

keys to change

 the value

Press ENTER 

key to store the value &

 move to next parameter

First Parameter

on Page-10

New Parameter 

value

Next Parameter

on Page-10

Main Display

Mode

Default Page

Page Number

Press PAGE 

key to enter

Set-up Mode

Use UP/DOWN 

key to set the 

Page Number

Press ENTER 

key to open

the Page

SET-UP MODE : ACCESS AND OPERATION

Notes

1.     To exit the set-up mode and return to the MAIN Display Mode, press and release PAGE key. 

2.     If no key is pressed for approximately 30 seconds, the set-up mode times out and reverts to the MAIN

 

Display Mode.

or

or

Section 3

 User Manual

Clave   

Содержание claveX 48 x 48

Страница 1: ...User Manual Autoclave Controller Clave...

Страница 2: ...ELLAYOUT 1 2 BASIC OPERATION 3 3 SET UPMODE ACCESS OPERATION 6 4 INPUT OUTPUT CONFIGURATION PARAMETERS 8 5 SOAK TIMER PARAMETERS OPERATION DETAILS 9 6 SUPERVISORYPARAMETERS 11 7 ELECTRICALCONNECTIONS...

Страница 3: ...or the balance Soak Time value depending on the autoclave cycle state In Program Mode the Lower Readout displays parameter names prompts INDICATORS TheTable 1 1 lists each front panel LED and the asso...

Страница 4: ...Pressing once decreases the value by one count keeping pressed speeds up the change PAGE User Manual Clave Table 1 1 Status Indicator HTR AIR Indicates Heater On Off status Indicates Air Outlet valve...

Страница 5: ...icating that the SP value is being shown on the Lower Readout Both the heater and air outlet are maintained ON As soon as the Temperature reaches theAir Outlet SP the air outlet is turned OFF Step 2 T...

Страница 6: ...e deviates above the control setpoint by more than the set band value see Operator Page and Parameters 3 PV Error The Alarm activates if the measured temperature value crosses the specified Max or Min...

Страница 7: ...e AIR OUTLET SETPOINT The setpoint value at and above which the Air Outlet valve is kept closed to build saturated steam pressure The Setpoint value at which the autoclave temperature value is maintai...

Страница 8: ...s to the MAIN Display Mode 4 Press and release the ENTER key until the prompt for the required parameter appears on the Lower Readout The last parameter in the list rolls back to the first parameter 5...

Страница 9: ...h the Lock enabled The Figure 3 2 below illustrates the Locking procedure Figure 3 2 UnLocking Repeat the Locking procedure twice for unlocking MASTER LOCKING The controller facilitates locking all th...

Страница 10: ...Control and Air Outlet setpoints HYSTERESIS Sets a differential dead band between the ON and OFF heater states Keep it large enough to avoid frequent switching of the heater without losing the desire...

Страница 11: ...1 to 5 0 Default 1 0 Re Start Abort Continue Default Continue Refer end of this section for detailed Soak Timer Operation Table 5 1 TIMER START BAND The temperature band around the control setpoint T...

Страница 12: ...on The timer is also provided with a Hold Band that can be enabled to make sure that the timer counts down only while the Temperature Value is within the Hold Band That is the timer pauses stops count...

Страница 13: ...esired rapid changes in the measured The higher the filter Temperature Value value the better the averaging but the slower the response to actual changes OFFSET FOR TEMPERATURE VALUE This parameter ad...

Страница 14: ...oth reading and writing No The Read Write parameters can only be accessed for reading That is the parameter values cannot be altered through serial communication Available for Utility Option Serial Co...

Страница 15: ...Use low resistance copper conductor leads of the same gauge and length Avoid joints in the cable Section 7 User Manual Clave 48 X 48 96 X 96 OUTPUT 1 Heater Control Output The Output 1 is factory conf...

Страница 16: ...ual Clave Line Neutral 12 L 13 N 2 Pole Isolating Switch Fuse Power Supply Terminal POWER SUPPLY Figure 7 6 Figure 7 4 a 18 17 16 NO C NC 16 17 18 48X48 96X96 As standard the controller accepts single...

Страница 17: ...a n e t Process Precision Instruments 101 Diamond Industrial Estate Navghar Vasai Road E Dist Palghar 401 210 Maharashtra India 0250 2391722 33 37 42 07498799226 09321985369 support ppiindia net sales...

Отзывы: