![PowerTek 9010 Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/powertek/9010/9010_user-manual_1591687066.webp)
Voltage range summary:
0.200000 V – 280.0000 V
Current range summary:
0.200000 mA – 20.50000 A
Electric power range:
0.04000 mW – 5.7400 kW
Frequency range:
DC, 15.000 Hz – 1000.00 Hz
Energy period range:
2.000 s – 3600.000 s
Phase shift range:
0.00° – 359.99°
Phase shift uncertainty:
0.15° at 200 Hz and below
0.25° above 200 Hz
Uncertainty of DC power is determined by formula
, where:
- dP is uncertainty of output power in %
- dU is uncertainty of set voltage in %, as per DCV specifications in chapter 8.1
- dI is uncertainty of set current in %, as per DCI specifications in chapter 8.2
This example calculates uncertainty of 4 W (20 V, 200 mA). Individual items in formula are calculated
like this:
- dU = 0.0035 % + 40 µV = 0.0037 % of 20 V value
- dI = 0.015 % + 6 µA = 0.018 % of 200 mA value
DC power uncertainty is then calculated as dP =
(0.0037
2
+ 0.018
2
+ 0.01
2
) = 0.021 %, which is also
the best DC accuracy achievable.
Uncertainty of AC power is determined by formula
, where:
- dP is uncertainty of output power in %
- dU is uncertainty of set voltage in %, as per ACV specifications in chapter 8.1
- dI is uncertainty of set current in %, as per ACI specifications in chapter 8.2
- dPF is uncertainty of power factor in %
dPF is based on type of power units used, calculated by following formulas:
-
for active power
-
for reactive power
-
for apparent power
where:
-
is set phase shift between voltage and current
- d
is uncertainty of set phase shift
dP =
(dU
2
+ dI
2
+ 0.01
2
)
dP =
(dU
2
+ dI
2
+ dPF
2
+ 0.03
2
)
dPF = [1 – cos (φ + dφ) / cos (φ)] ∙ 100
dPF = [1 – sin (φ + dφ) / sin (φ)] ∙ 100
dPF = 0
Powertek MC9010 User Manual
Содержание 9010
Страница 1: ......
Страница 2: ......
Страница 3: ...Powertek MC9010 User Manual ...
Страница 4: ...Powertek MC9010 User Manual ...
Страница 5: ...Powertek MC9010 User Manual ...
Страница 32: ...Following items can be modified Powertek MC9010 User Manual ...
Страница 45: ...Powertek MC9010 User Manual ...
Страница 46: ...Powertek MC9010 User Manual ...
Страница 47: ...Powertek MC9010 User Manual ...
Страница 48: ...Powertek MC9010 User Manual ...
Страница 49: ...Powertek MC9010 User Manual ...
Страница 53: ...Powertek MC9010 User Manual ...
Страница 55: ...Powertek MC9010 User Manual ...
Страница 74: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual Powertek MC9010 User Manual ...