Adjustment is done through MENU > Calibration > Data. This menu item is password protected, default
factory set calibration code is “9010”.
Calibration data are sorted in a tree structure with following hierarchy:
1. Function groups (source and measurement)
2. Functions (f.e. Voltage DC)
3. Ranges (f.e. 10 V)
4. Calibration points (f.e. zero offset)
As with any other menu you can navigate through the tree structure using cursor buttons, rotary knob or
softkeys. symbol indicates calibration tree branches with out-of-date or otherwise invalid calibration
data:
Powertek MC9010 User Manual