Select SCOPE function using SCOPE key, following screen will appear:
Push the SCOPE key repeatedly to swap between Voltage – SINE – PWM – Time marker mode.
DC and squarewave calibrated signals are available in this mode. Requested mode Voltage can be
selected after pushing soft key Shape. Following list of waveforms is displayed:
- DC Positive
DC positive voltage with adjustable level
- DC Negative
DC negative voltage with adjustable level
- Positive square
Positive square wave signal with adjustable duty cycle ratio and frequency
- Negative square
Negative square wave signal with adjustable duty cycle ratio and frequency
- Symmetric square Symmetric square wave signal with adjustable duty cycle ratio and frequency
Voltage mode has two submodes Low and High. In Low mode output signal is connected to the front
panel coaxial connector. This mode offers frequency range to 100 kHz for square wave signals.
In High mode output signal is connected to HI – LO output terminals. It is identical to ACV function.
Displayed units of amplitude can be set by pressing
SELECT key to enter main value edit mode, then
pressing Unit softkey and choosing one of the following units: V
rms
(root mean square), V
pk
(peak), V
pp
(peak – peak), V
avg
(average), dBu (dB microvolt), dBV (dB volt), dBm (dB miliwatt).
To calibrate an oscilloscope:
1. Connect the oscilloscope to N connector (Low mode) or to HI – LO terminals (Low) mode.
2. Set appropriate mode Low/High using Mode softkey.
3. Select requested waveform using soft key Shape and confirm with Enter.
4. Set requested frequency, amplitude, duty cycle ratio.
5. Push the OPER key to connect signal to the output terminals.
Sine mode offers calibrated sinewave voltage with low distortion and stabilized amplitude. Output
impedance is 50Ω. Select the Sine mode by pushing the SCOPE key until screen with Sine mode is
displayed. Following values are available for setting:
- Frequency in Hz
- Amplitude in preselected unit Vrms, Vpk, Vpp, Vavg, dBµ, dBV, dBm
Push the OPER key to connect signal to the output terminals (N connector).
Powertek MC9010 User Manual
Содержание 9010
Страница 1: ......
Страница 2: ......
Страница 3: ...Powertek MC9010 User Manual ...
Страница 4: ...Powertek MC9010 User Manual ...
Страница 5: ...Powertek MC9010 User Manual ...
Страница 32: ...Following items can be modified Powertek MC9010 User Manual ...
Страница 45: ...Powertek MC9010 User Manual ...
Страница 46: ...Powertek MC9010 User Manual ...
Страница 47: ...Powertek MC9010 User Manual ...
Страница 48: ...Powertek MC9010 User Manual ...
Страница 49: ...Powertek MC9010 User Manual ...
Страница 53: ...Powertek MC9010 User Manual ...
Страница 55: ...Powertek MC9010 User Manual ...
Страница 74: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual Powertek MC9010 User Manual ...