FirstCall™ Liq x Liq Medical Gas Manifold
Operating & Maintenance Manual
Powerex • 150 Production Drive • Harrison, OH 45030 • USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
IN599800AV • 15 January 2021
Page 10 of 24
For user select “USER.”
Default password is
Pass123
(case sensitive). This
password can be changed in the settings screen.
Once the correct password is entered, the Manifold
will now be in “Settings” mode. A “Settings” gear icon will
appear at the bottom.
Press the “Settings” icon to get to the Settings Menu.
Alarm History
The Alarm History screen shows a record of all alarm
events up to 32, including the specific alarm condition
and the date and time of the alarm event.
This screen also has the ability to sort the alarm history
list by a number of different criteria. The default sort is
“Time”. The alarm history list can be sorted differently
by pressing the Alarm History Sort button in the upper
left corner.
Alarm Status
The Alarm Status screen shows a list of all possible
alarm conditions for the maximum number of gases
and input signals available in the Powerex FirstCall ™
Alarm System.
The default alarms for each manifold are as follows:
• Main line pressure high
• Main line pressure low
• Changeover to secondary supply
• Emergency reserve in use
• Emergency reserve low