Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product 
described. Protect yourself and others by observing all safety information. Failure to comply with instructions could result in 
personal injury and/or property damage! Retain instructions for future reference.

Powerex • 150 Production Drive • Harrison, OH  45030 • USA
1-888-769-7979 • www.powerexinc.com

Claw Medical Vacuum Systems

IN561501AV 5/12

Description

General

Powerex Medical Vacuum Systems are designed to provide 
vacuum for hospitals and medical institutes. They are 
designed and constructed to meet NFPA 99 requirements 
for Level 1 piped vacuum systems. If WAGD vacuum is the 
same as the medical-surgical vacuum source: Flammable 
anesthetics or flammable vapors shall be diluted below the 
flammable limit prior to disposal into the vacuum system 
or the pumps shall be designed of materials and using 
lubricants that are inert in the presence of oxygen, nitrous 
oxide, and halogenated anesthetics.

Safety Guidelines

This manual contains information that is very important 
to know and understand. This information is provided for 
SAFETY and to PREVENT EQUIPMENT PROBLEMS. To help 
recognize this information, observe the following symbols.

 

Danger indicates an imminently  

  hazardous situation which, if 

not avoided, WILL result in death or serious injury.
 

Warning indicates a potentially  

  hazardous situation which, if 

not avoided, COULD result in death or serious injury.
 

Caution indicates a potentially 

  hazardous situation which, if 

not avoided, MAY result in minor or moderate injury.
 

Notice indicates important  

  information, that if not 

followed, may cause damage to equipment.

Unpacking

Immediately upon receipt of the vacuum system, inspect 
for any damage which may have occurred during shipment. 
Repair or replace damaged items before use. The name-plate 
should be checked to verify the correct model and voltage.

 

 

Do not operate unit if damaged  

  during shipping, handling or 

use. Damage may result in unsafe conditions and cause injury 
or property damage.

Specifications

Product

CVTD, CVPD, CVPT, CVPQ Series

Operating 
Voltages

208V, 230V, 460V

Control Panel

UL508A and NFPA compliant

Motor

TEFC Electric Motor

Tank

ASME Rated for 200 psi MAWP

Pump

Capable of operating at:
28.4” Hg (sea level) continuous duty for 2 HP
28.4” Hg (sea level) continuous duty for 3 HP 
28.4” Hg (sea level) continuous duty for 5 HP
28.4” Hg (sea level) continuous duty for 6.4 HP
27.0” Hg (sea level) continuous duty for 7 HP
27.0” Hg (sea level) continuous duty for 9.1 HP
25.5” Hg (sea level) continuous duty for 10.2 HP
23.4” Hg (sea level) continuous duty for 15 HP

Drive

Direct

Tank Sizes

80 Gallon to 240 Gallon

Performance

See Cut Sheet

Outlet Air 
Connections

2 HP - 1 inch
3 HP - 1 inch
5 HP - 1 inch
6.4 HP - 1 inch
7 HP - 1 inch
9.1 HP - 1 inch
10.2 HP - 1-1/2 inch
15 HP - 3 inch

Содержание CVPT Series

Страница 1: ...not avoided WILL result in death or serious injury Warning indicates a potentially hazardous situation which if not avoided COULD result in death or serious injury Caution indicates a potentially haz...

Страница 2: ...eiver Tank The receiver tank in a Powerex Medical Vacuum System is plumbed so that it sees the system vacuum It is not pressurized in normal use and should not be connected to any pressure sources The...

Страница 3: ...ctrical supply before servicing the equipment Electrical shock hazard Make sure the system is grounded in accordance with NEC and local requirements All electrical hook ups must be performed by a qual...

Страница 4: ...f the system extends beyond 600 equivalent straight feet use one size larger pipe Approximate system CFM equals the number of pumps running times CFM in table below Data below is for reference only if...

Страница 5: ...he vacuum level is maintained and the minimum run timer conditions are satisfied 7 Check for excessive vibration unusual noises or leaks during operation If problems are detected shut down the system...

Страница 6: ...maintenance schedule relates to the normal operating conditions If the circumstances and load condition are adverse shorten the interval time and perform maintenance accordingly Maintenance LUBRICATIO...

Страница 7: ...eck drive coupling replace if needed 2 Verify that pump shaft turns freely repair or replace as necessary 3 Check vacuum discharge piping to verify air is being exhausted repair or replace as necessar...

Страница 8: ...ns iii that have been materially modified without the prior written approval of Powerex or iv that experience failures resulting from operation either intentional or otherwise in excess of rated capac...

Отзывы: