Sitter On Cue PRO
18
Monitoring with Multiple Sensors
Up to five sensors (2 wired and 3 wireless) can be connected to the fall monitor allowing
the patient to be transferred from one location to another without moving sensors or
switching out cords.
NOTE:
Only one wireless sensor of each type can be paired at a given time (chair, bed
or toilet). While the fall monitor is NOT actively monitoring, if a second wireless sensor of
the same type is paired, it will replace the existing sensor. When the fall monitor is actively
monitoring, an audio cue will indicate that the new sensor could not be paired. Press
HOLD to pause monitoring before connecting a new wireless sensor of the same type.
Transferring a patient between sensors
1.
Connect the second sensor to the fall monitor (First sensor may be plugged in or paired, if
wireless).
2.
Ensure fall monitor is on.
3.
Press HOLD button to pause monitoring for 30 seconds. You have 30 seconds to remove the
patient from the current sensor pad or unfasten the sensor belt. During this time the STATUS LED
will flash red.
4.
After 30 seconds, and the patient is off the sensor pad, the alarm will remain in the Ready state.
5.
Place the patient on the second sensor or fasten the sensor belt. If the patient is placed on
the sensor pad before the hold expires, the HOLD button can be pressed again to resume
monitoring.
NOTE:
While multiple sensors can be connected to the fall monitor, only one can be used
for monitoring at a time. If a first sensor is actively monitoring and pressure is applied to a
second sensor or the sensor belt is fastened, an audible cue will indicate a second sensor
is about to be activated and gives 10 seconds for the person to remove pressure from or
unfasten the sensor. If pressure is not removed or the sensor belt is not unfastened, the
alarm will sound. Press the HOLD button. You have 30 seconds to remove pressure from
at least one of the sensor pads or unfasten the sensor belt.
MONITORING