Sitter On Cue PRO
16
Latching Mode
In non-latching mode (factory default), when weight is re-applied to a sensor or the sensor belt is
reattached the Sitter On Cue PRO will stop signaling the nurse call system.
When Latching mode is selected and an alarm signal is initiated, the nurse call signal will cease
only when the caregiver interacts with the alarm. Even after the patient returns to the sensor pad,
the caregiver must press the HOLD button to stop the alarm from signaling the nurse call system.
Configuring Latching Mode:
1.
Access Advanced Setup Mode by pressing and holding the HOLD button while moving the power
switch from the OFF to the ON position.
2.
Rapidly press the hold button three times to toggle between Latching and Non-Latching mode.
3.
With Latching Mode selected, press the tone button to switch between the following two options:
•
“Alarm ceases when patient returns.” – In this mode, if the patient applies pressure to the
sensor or reattaches the sensor belt, the alarm tone will stop. The audio cue “Press hold to
reset alarm” will repeat every 15 seconds until the HOLD button is pressed. This is the factory
default setting.
•
“Alarm continues when patient returns.” – In this mode, the alarm tone will continue until the
HOLD button is pressed.
LATCHING MODE