33 

Troubleshooting Chart 

 

Problem 

Possible Causes 

Corrective Action 

Unit does not run 
(Digital Display is 
blank) 

No power to unit 
 

Refrigerated units only 

Electrical cord between 
Control Head and 
Refrigeration Power Module 
loose or unplugged   
Power Switch / Circuit 
Breaker in OFF position

 

Check that the electrical cord is secure and 
connected to an operating electrical outlet. 
 
Check that the electrical cord is secure and 
properly connected. 
 
 
Place Power Switch / Circuit Breaker in ON 
position. 

Unit does not run 
(STANDBY appears 
on Digital Display) 

Unit in Standby mode 

Press Power Key on front panel. 

No fluid circulation 

Insufficient fluid in reservoir 
Pump impeller jammed 

Add fluid to reservoir. 
Inspect pump and remove debris as required. 

Insufficient circulation 

Fluid viscosity too high 
External tubing diameter 
too small 
Low line voltage 

Replace with lower viscosity bath fluid. 
Replace with larger diameter tubing. 
 
Check and correct as required. 

Unit does not heat 

Insufficient fluid in reservoir 
Temperature set point too 
low 
Safety Set Temperature too 
low 

Add fluid to reservoir. 
Increase temperature set point. 
 
Increase Safety Set temperature. 

Insufficient heating 

Insufficient circulation 
Low line voltage 
Ambient temperature too 
cool 
Excessive heat loss 

See Insufficient circulation, above. 
Check and correct as required. 
Increase ambient temperature or relocate unit. 
 
Check for heat loss from external tanks and 
hoses; 
Check for vapor/heat loss from internal 
reservoir. 

Temperature 
unstable 

Insufficient circulation 
Debris or mineral build-up 
on pump, heater, or 
temperature sensor. 

Check pump flow and operation.  
Clean as required. 

Unit does not cool 

Dust build up on air filter or 
condenser 
Blocked air ventilation 
screens 
Temperature set point is 
too high 
Excessive heat load 
 
Ambient air temperature too 
high (>35°C / 95°F) 
Low or high line voltage 

Clean air filter and/or condenser as required. 
 
Remove blockages as required. 
 
Decrease temperature set point. 
 
Check that heat load does not exceed capacity 
of bath; correct as required. 
Decrease ambient air temperature. 
 
Check and correct as required. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание MX06S135

Страница 1: ...07 May 2018 Models MX07R 20 MX7LR 20 MX15R 30 MX20R 30 MX07H135 MX15H135 MX20H135 MX06S135 MX10S135 MX20S135 MX28S135 MX08P100 MX11P100 MX14P100 MX17P100 MX23P100 MX28P100 MX17VB6G MX27VB6G MX28C135 w...

Страница 2: ...et Calibration 21 Setting the Low Limit Temperature 22 Setting the High Limit Temperature 23 Adjusting the Auto Cool Temperature 24 Resetting the Factory Default Values 25 Loss of Power Restart 25 Tap...

Страница 3: ...eating Bath 7 liters 20 to 135 C 4 to 275 F MX15R 30 Refrigerating Heating Bath 15 liters 30 to 135 C 22 to 275 F MX20R 30 Refrigerating Heating Bath 20 liters 30 to 135 C 22 to 275 F MX07H135 Heating...

Страница 4: ...electricity or electric shock This symbol indicates that a hot surface may be present This symbol marks information that is particularly important This symbol indicates alternating current These symb...

Страница 5: ...fluid in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptable levels Always refill the reservoir using the same bath fluid that is already in the reservo...

Страница 6: ...mpatibility Directive 2004 108 EC IEC 61010 1 2001 IEC 61010 2 2001 IEC 61326 2005 EN 61326 2006 Unpacking Your Circulator Your Circulator is packed in a special carton or cartons You should keep the...

Страница 7: ...IEC Power Cord 6 ft 1 82 m IEC to Mains Power Cord Refrigeration Control Cable Fitting Kit for External Applications Cooling Coil N A Integral Optional 1 Optional 2 Optional 2 Qualification Test Resu...

Страница 8: ...cess panel Drain Valve and Port right side on MX7LR 20 only Backlit Digital Display Up Arrow Button Down Arrow Button Menu Button Power Button Safety Thermostat under cover plate Actual Bath Temperatu...

Страница 9: ...Always disconnect electrical power and drain fluid from bath before moving WARNING To avoid the potential for burns allow the Circulator to cool completely before cleaning or performing any maintenanc...

Страница 10: ...ARNING The top deck on Viscosity Baths is not attached Do not remove deck while Circulator is operating Do not lift bath by grasping the Temperature Controller or top deck Always disconnect electrical...

Страница 11: ...110 509 PSC EN 10 Coliform Bath MX Controller Hinged Gable Cover Lidded Port for Optional Cooling Coil www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 12: ...ker ON Refrigerating Heating Baths only Maximum 1 in 2 54 cm below the underside of the top deck Minimum 2 6 in 6 7 cm below the underside of the top deck Heat only Circulators IEC Power Cord Temperat...

Страница 13: ...0 509 PSC EN 12 4 Turn Controller ON 5 Enter temperature set point 6 Set safety thermostat once unit reaches set point www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply c...

Страница 14: ...ing reservoir WARNING See Technical Information in the rear of this manual for a list of compatible liquids WARNING If the proper fluid level is not maintained the heater coil may become exposed and p...

Страница 15: ...ts on the rear of the MX Controller To maintain adequate flow avoid restrictions in the tubing If a higher flow rate is desired you can move the flow adjuster toward a closed position as shown below T...

Страница 16: ...e Refrigeration Control Cable connected to the rear of the Temperature Controller to the Refrigeration Control Connection on the Refrigeration Power Module Refrigeration Control Cable Refrigeration Co...

Страница 17: ...ower Module Attach the 6 ft 1 8 m power cord to the IEC electrical connection on the Refrigeration Power Module and then plug the male connector into the Mains electrical outlet Place the Power Switch...

Страница 18: ...ll begin flashing Continue pressing the and keys until the set point temperature is equal to your desired Safety Set temperature SET will stop flashing approximately 10 seconds after the desired tempe...

Страница 19: ...ll characters symbols on the LCD will momentarily light When the Circulator begins running the actual and set point temperatures will be displayed and the circulating symbol will be lit If the actual...

Страница 20: ...value in a sub menu press or allow the display to time out approximately 10 seconds Adjusting the Temperature Set Point This is the temperature at which the fluid in your Circulating Bath will be mai...

Страница 21: ...mperature are displayed The factory default is C To Access Press the key until C F appears on the display To Change Press to select C press to select F To Accept Press or allow the LCD to return to th...

Страница 22: ...e power cord from the electrical outlet 2 Plug the power cord back into the electrical outlet while pressing and holding the key 3 When STANDBY appears on the display release the key and press You can...

Страница 23: ...to 50 F the factory default is 35 C 30 F To avoid unwanted alarms or shutdowns during regular operation the Low Limit value should be set at least 5 lower than the selected control temperature To Acce...

Страница 24: ...100 to 285 F the factory default is 140 C 285 F To avoid an unwanted shutdown during regular operation the High Limit value should be set at least 5 higher than the selected control temperature To Acc...

Страница 25: ...ontrol system Cool Command allows the refrigeration system to turn on at a fluid temperature up to 135 C when the temperature set point is changed to or below the Auto Cool set point 135 C maximum As...

Страница 26: ...lue Temperature Scale C Temperature Set Point 20 C 68 F Calibration Offset 0 0 C Low Limit 35 C 30 F High Limit 140 C 285 F Auto Cool 45 C Loss of Power Restart WARNING The unit will start automatical...

Страница 27: ...se on the Coliform Bath and Stainless Steel Open Bath Systems This cooling coil has two 0 375 inch 9 5 mm OD straight barb fittings at which the fluid inlet and outlet connections can be made Either c...

Страница 28: ...em is corrected Allow bath to warm or add heat load Decrease the Low Limit temperature value Alarm The bath temperature has risen above the High Limit temperature value Power to the heater and pump wi...

Страница 29: ...e reservoir an algaecide such as polyclean Algaecide 004 300040 should be used WARNING Do not use chlorine bleach Draining the Bath Reservoir WARNING Bath fluids should be stored and disposed of accor...

Страница 30: ...t of approximately 50 C and then allow the bath to stabilize at that temperature The amount of time this will take will depend on the size of the bath and the difference between the initial bath tempe...

Страница 31: ...etted parts The cleaner should be added to the bath reservoir at the prescribed dosage and circulated at 60 C 140 F until the scale is removed External Surfaces Only mild detergents and water or an ap...

Страница 32: ...al tools It is removed as follows 1 Place the tip of a small flat blade screwdriver under the retaining ring locking tab and pry up gently 2 Rotate the Temperature Controller clockwise until it stops...

Страница 33: ...oriented above one of the indents on the top deck slowly lower the Temperature Controller into the top deck opening until it is resting on top of the positioning pins 2 Gently rotate the Temperature C...

Страница 34: ...luid in reservoir Temperature set point too low Safety Set Temperature too low Add fluid to reservoir Increase temperature set point Increase Safety Set temperature Insufficient heating Insufficient c...

Страница 35: ...low end extreme temperatures Pump speed too high Incorrect bath fluid Insufficient insulation on external fluid lines Ambient air temperature too high 35 C 95 F Low or high line voltage Dust build up...

Страница 36: ...135 C Ambient 20 to 275 F 120V 60Hz 10A 240V 50Hz 6A MX06S135 Stainless Steel Open Tank Bath 6 liters Ambient 10 to 135 C Ambient 20 to 275 F 120V 60Hz 10A 240V 50Hz 6A MX10S135 Stainless Steel Open...

Страница 37: ...10 to 135 C Ambient 20 to 275 F 120V 60Hz 10A 240V 50Hz 6A 1 Maximum operating temperature for polycarbonate tank MX Controller is capable of higher temperatures Environmental Conditions Indoor use on...

Страница 38: ...e You are responsible for proper selection and use of the fluids Avoid extreme range operation Fluid Description Viscosity cSt 25 C Specific Heat Normal Temperature Range Extreme Temperature Range Flu...

Страница 39: ...y affects temperature stability A viscosity above 10 centistokes adversely affects temperature uniformity A high fluid viscosity and high pump speed adds heat to the fluid being pumped At a fluid s te...

Страница 40: ...It is also your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect the persons involved in the disposal and recycling of th...

Страница 41: ...8L Polycarbonate Open Bath 510 698 Bath Cover for 6L Stainless Steel Open Bath 510 700 Bath Cover for 10L Stainless Steel Open Bath 510 701 Bath Cover for 20 and 28L Stainless Steel Open Baths 510 702...

Страница 42: ...r polyclean Algaecide 8 oz 236 ml 004 300040 polyclean Algaecide Twelve 8 oz 236 ml bottles 004 300041 polyclean Bath Cleaner 8 oz 236 ml 004 300050 polyclean Bath Cleaner Twelve 8 oz 236 ml bottles 0...

Страница 43: ...t should be shipped transportation charges pre paid in either its original packaging or packaging affording an equal degree of protection to the indicated service center To insure prompt handling the...

Отзывы: