Setting the High Limit Temperature
This sub-menu allows you to limit how high the temperature set point may be set. It also serves as a high
limit safety, alerting you if bath temperature rises above the high limit temperature setting. The High Limit
value may be set from +40° to +140°C / +100° to +285°F; the factory default is +140°C / +285°F.
To avoid an unwanted shutdown during regular operation, the High Limit value should be set at least 5°
higher than the selected control temperature.
To Access:
Press the
key until HIGH LIMIT appears on the display.
To Change:
Press
or
.
To Accept:
Press
or allow the LCD to return to the main operational display
(approximately 10 seconds).