7
【
CONTROL PANEL
】
1.
LED Digital Display:
show passer number.
2. The alarm number.
3.
Interference Signal Strength Indicator:
alarm indicator (red light)
4.
“ENTER" Key:
confirm the data after setting, and use as plus key when setting
the district sensitivity.
5.
“SETUP" Key:
①
convert various data, and use as minus key when setting the district sensitivity.
②
long press 5s to adjust working frequency, 1~16 bands to choose.
6.
"SELECT" Key:
choose each program, and use as plus key when entering
password .
7.
"RESET" Key:
zero clearing for number of passer and alarm.
【
DETECTION ZONES
】
A: Control Unit: signal strength indicator, the number of passer and alarm
Standby Indication: indicator in red when alarming.
B: LED bar Alarm
As Figure-1, there are six sets of unique pinpointing LEDs, which represent six
detection zones(from bottom to top 1.2.3.4.5.6).Same way for Figure-2 and Figure-3.
Pinpointing LEDs have 2 states in red bright or not bright. It’s in red alarm when
passers’ body with or exceed stated metal content,if there are multiple targets, then
each target location of the LEDs will be lit with alarm sound at the same time, (Except
for mute).
C: Infrared Sensor(IR)
When the power is turned on, the WTMD starts to work, the infrared sensor can
effectively stop the alarm when nothing pass it to avoid false alarm. The passers could
be exactly counted by IR. Regardless of any cause of the alarm in transit, he has to go
through the WTMD again.