18 Zone Walk Through Metal Detector
1. Long press “SELECT”3 seconds, entering“P-0”or ”P-1”mode. (“P-0” for
Inspection, “P-1” for Test Flying Object or Test Mode.) Adjust to “P-0” mode and
press “ENTER”to confirm entering P-0 mode.
2. Long press“SETUP” entering“ Frequency Adjustment”,will show “F-xx”.
“xx”means Device Frequency Code. Press “SELECT” can adjust the Frequency. Press
“ENTER” confirm the selection. In order to avoid signal interference, the frequency
code of each adjacent device must be different. (For user convenient operation, we
have particularly simplified the Frequencies to 16 total, which can for unlimited
devices installation!)
3. Press “SELECT”, will show “12345”, which is factory original password. If have
changed the original password, please input the new correct password. (For password
input details please refer to “Password Management”)
4. Press “ENTER” again, will show “00000”, means password correct can go to next
setting. Or if show “E0000”, means password wrong, need input the correct password
then can go to the next setting.
5. One more time press “ENTER”, will show “A-18”, means the device of 18 zones.
6. Press “SETUP”, will show “d-01”, which is alarm duration time. “01”means alarm
duration of 1 second, maximum alarm duration of 3 seconds. Press “SETUP” can
adjust to 01, 02, 03 second.
7. Press “SELECT” entering Zone-Sensitivity-Level adjustment, will show “1L-xxx”.
( “1” means 1st zone, “L”means left side zones, “xxx” means zone sensitivity level.)
Press “SETUP” can reduce the zone sensitivity level, press “ENTER” can increase the
zone sensitivity level. The higher the “xxx”figure, the higher the sensitivity!
Example, adjust to “1L-365”, means the Left side 1st zone sensitivity level is 365, the
detectable metal object would be a coin or a smaller metal object.
Samely, press “SELECT” in turn, will show “2L-xxx”~“6L-xxx”, user can adjust the
left side 6 zone sensitivity level respectively.
8. Then press “SELECT”again, entering right side zone settings. Will show “1R-xxx”,
means right side first zone sensitivity level. Same as the “1L-XXX” zone sensitivity
adjustment, user can adjust the right side 6 zone sensitivity level respectively.
9. Now “Left” & “Right”side Zone-Sensitivity-Level finish adjusting, press
“SELECT” entering whole Sensitivity-Level adjustment, will show “H-xxx”.