COMBINED CHEMICAL AGENT  

AND GAMMA RADIATION DETECTOR PM2012M 

 

Models 

PM 2012M 

PM 2012MA 
PM 2012M

В 

 

Operation Manual 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание PM 2012M

Страница 1: ...COMBINED CHEMICAL AGENT AND GAMMA RADIATION DETECTOR PM2012M Models PM 2012M PM 2012MA PM 2012M Operation Manual...

Страница 2: ...ates endeavor to assure that their published engineering specifications and manuals are correct Despite these efforts errors can occur Polimaster Ltd and its affiliates reserve the right to correct an...

Страница 3: ...ound the world that have recognized the capability and quality of Polimaster s instruments to assist them in their daily detection operations The PM2012M PM2012MA PM2012MB are ideally suited for a lar...

Страница 4: ...INSTALLATION OF USE OF AND RESULTS OBTAINED FROM THE PRODUCT MANUFACTURER DOES NOT WARRANT THAT THE PRODUCTS AND SERVICES HEREUNDER WILL MEET PURCHASER S OR USERS REQUIREMENTS OR WILL OPERATE IN THE C...

Страница 5: ...S HAVE BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES Limitation on Liability Notwithstanding any provision to the contrary herein and with the exception of the claims of intellectual property rights...

Страница 6: ...Detector Module 19 3 4 2 Checking the CWA Module 19 Chapter 4 Operation Guidelines 22 4 1 Introduction 22 4 2 Dose rate measurements 23 4 3 Dose measurements 24 4 4 Detection of the chemical warfare...

Страница 7: ...assembling of the chemical agent detection module and removal of the radioactive source will result in radiation exposure To avoid radiation exposure never disassemble the chemical agent detection mod...

Страница 8: ...strictly observed Do not operate the unit if it is damaged or damage is suspected Pack the unit in its original packing case and return to the manufacturer or authorized service center for servicing...

Страница 9: ...transfer via USB with PC and via Bluetooth channel with smartphones The unit also detects vapors emitted from chemical warfare agents CWA organophosphorus compounds sarin soman V gases and arsenic co...

Страница 10: ...dose equivalent for the PM2012M model for the PM2012MA PM2012MB models from 1 Sv to 9 99 Sv from 1 Sv to 14 9 Sv Main relative dose measurement error 15 Dose accumulation time up to 99 999 h PM2012M P...

Страница 11: ...discharge of 8 kV contact discharge of 6 kV test level 3 Thresholds of detection of organophosphorus compounds arsenic containing compound 5 1 5 10 5 mg l detection time is less than 10 s 3 0 9 10 4...

Страница 12: ...nents The PM2012M PM2012MA PM2012MB GP Alkaline LR20 size D Battery CWA simulator Measuring cup Fabric filter elements 1 and 2 white Fabric filter element black Bottle with filter filler Cleaning rod...

Страница 13: ...13 OPTIONAL ACCESSORIES 220V 12V AC adapter and 9 36 V DC power cable Rotameter Infrared adapter ACT IR220L or IR 210B USB cable for PM2012MB...

Страница 14: ...WA Module The analyzed air enters the unit via the inlet 1 and exits it via the outlet 11 The display 6 together with two color LEDs bars 3 and 3 along with the audible alert device 4 provides all rel...

Страница 15: ...ion The Radiation Detector module continuously measures the ambient dose equivalent rate 10 and the ambient dose equivalent 10 of gamma and X ray radiation and the dose accumulation time If the preset...

Страница 16: ...onitors the battery voltage and displays the battery sign with the analogue scale in the upper right corner If the battery is fully charged all five marks are displayed In case of critical discharge t...

Страница 17: ...nnector is blue connector is To avoid spontaneous switching the CWA operating modes do not expose the instrument to strong constant magnetic fields 3 3 Instrument Turning On Off The instrument turns o...

Страница 18: ...ts measurements but the CWA module is still not working To operate with the CWA module three control marks on the rotating head 2 Figure 2 1 are used module is switched off calibration mode CWA detect...

Страница 19: ...d dose rate value exceeds the preset threshold an audible alarm will sound Remove the source The audible alarm should terminate in 30 seconds Figure 3 5 The mark of the geometrical center of the gamma...

Страница 20: ...his mode ensures the additional blowing of the ionization chamber The instrument automatically returns to the calibration mode when the blowing is successfully completed CAUTION If the instrument cont...

Страница 21: ...t 1 Figure 2 3 of the instrument and open it Red or yellow LEDs will blink L bar or GV bar respectively depending on the simulator type marked on its ends and the audible alarm will sound refer to sec...

Страница 22: ...DE button Figure 3 2 can switch the unit to the PC communication mode and see the current time hours minutes and seconds Dose equivalent rate screen Dose equivalent screen Time screen PC communication...

Страница 23: ...e screen Dose equivalent rate threshold screen Figure 4 2 The dose equivalent rate DER and DER threshold screens The manufacturer presets the DER threshold value at 1 Sv h The DER threshold setting ra...

Страница 24: ...igure 3 2 To return to the dose indication press the SET button again Dose equivalent screen Dose accumulation time screen Dose equivalent threshold screen Figure 4 3 The dose equivalent DE dose accum...

Страница 25: ...procedure the unit displays CHEM SEARCH ON refer to section 3 4 2 1 and Figure 3 7 switch the unit to the CWA detection mode turn the rotating head in the direction shown by the arrow to math the mar...

Страница 26: ...tected concentration is higher than about 125 of the preset threshold value Triple discontinuous audible signal with 1 s interval The manufacturer sets the CWA vapor detection thresholds as follow for...

Страница 27: ...requirements IBM PC Pentium III or higher 1 G free HDD space operating systems Microsoft Windows 7 Windows 8 or Windows 10 To use the PC communication mode install PM2012M Data Processing Software ref...

Страница 28: ...ment non volatile memory instrument operation history DER values in set time intervals time and date when the instrument was switched on off time date and current DER value when the preset DER thresho...

Страница 29: ...tion push the Home button iPhone and Android and switch off the Bluetooth channel as in par 2 4 The following operations can be conducted when the instrument is connected to smarthone in particluar to...

Страница 30: ...e maintenance procedures Protect the instrument from shocks and mechanical impacts corrosive environments organic solvents and open fire 5 1 Maintenance schedule Task Frequency Refer To Cleaning of th...

Страница 31: ...the rotating head 2 Figure 5 1 and the mark on the unit case align the unit vertically so that the inlet 1 Figure 3 2 is facing upward take the cleaning rod and a strip of cotton cloth 8 10 mm wide an...

Страница 32: ...3 2 is facing upward take a medicine or measuring pipette with 10 15 drops 0 5 ml of rectified ethanol insert pipette into the inlet not deeper than 1 2 cm and put in ethanol drop by drop into the io...

Страница 33: ...st of the micropump capacity Connect the lower silicone tube of the rotameter available as an option with the outlet 11 Figure 3 2 check that a seal is made tight Switch the instrument into the CWA mo...

Страница 34: ...ded in the Accessories kit 10 fabric filter element 2 white included in the Accessories kit Figure 5 1 Filter construction The replacement procedure consists of the following steps turn the instrument...

Страница 35: ...rough remove the fabric filter elements 9 and 10 install the new fabric filter elements 9 and 10 from the Accessories kit and the cleaned gauze 8 fill in the unit case with a fresh coal Figure 5 2 usi...

Страница 36: ...d be fulfilled by the specially trained personnel at the manufacturer facility or authorized service center 5 6 Instrument decontamination The protective treatment includes external examination of the...

Страница 37: ...isplays error messages Test failure Send the instruments to the manufacturer or authorized service center Instrument produces neither audible nor visual alarms in the CWA detection mode if the simulat...