26
Instructions for Use 2019-03-20
Other Components: Wiping with Ethanol (70 vol% to 80 vol%)
Disinfection
Components maintained this way:
Maintenance Procedure
Main Unit
Probe Cord
Tester
Disinfection
Wipe the components with a piece of gauze that has been
dampened with ethanol (70 vol% to 80 vol%) and wrung out
thoroughly.
• Do not use anything except ethanol (70 vol% to 80 vol%).
• Do not immerse the components in or wipe it with any of the following: functional water (acidic electrolyzed water, strong
alkaline solution, or ozone water), medical agents (glutaral, etc.), or any other special types of water or commercial clean-
ing liquids. Such liquids may result in metal corrosion and adhesion of the residual medical agent to the components.
• Never clean the components with chemicals such as formalin cresol (FC) and sodium hypochlorite. These will damage the
plastic parts of the components. If any of these liquids being applied to the components, wash it off in running water.
• Use only ethanol (70 vol% to 80 vol%) and OPTI-CIDE-3™ Surface Wipes for cleaning. Any other cleaning chemical or
products should not be used including but not limited to the following cleaning products and similar cleaning products
listed below because of the potential damage to the plastic components of the Endostar Navigator.
• CaviWipes™ • CaviCide™ • SANI-CLOTH™