Autoclavable Components
Cleaning
1. Disconnect the file holder, long file holder and contrary elec-
trode from the probe cord.
2. Clean them off in running water with a soft brush and then
wipe off the water.
If a medical agent being used for the treatment has adhered to
the components, wash it off in running water.
Do not clean the components ultrasonically.
After washing is complete, check to see if the file holder or
long file holder, including its inside, is completely dry. If any
water remains inside the component, expel it with an air gun
or another such tool. Failure to do so could result in the re-
maining water coming out during use and cause malfunction
or poor sterilization.
If dust or other impurities adhere to the hook of the file holder
or long file holder, they may cause malfunction.
Do not use the high-temperature washer-disinfector.
Hook