Autoclavable Components
Sterilization
Main Unit
Not Autoclavable
Autoclave the file holder, contrary electrode, and long file holder after use for
each patient.
Recommended temperature and time:
In a sterilization pouch, minimum 6 minutes at +134°C (+273.2°F) or minimum 60 minutes
at +121°C (+249.8°F).
Minimum drying time after sterilization:
10 minutes.
Never autoclave the main unit.
Autoclaving and drying temperatures must never 135°C (+275°F). Excess temperature could cause the
components to malfunction or could cause discoloration.
Take the file out of the file holder or the long file holder before autoclaving.
Clean everything throughly before autoclaving. Any chemicals or foreign debris left on components could cause
them to malfunction or could cause discoloration.
Do not leave the file holder, long file holder, and contrary electrode in the autoclave.
For sterilizing files, follow the manufacturer’s recommendations.
• To prevent the spread of serious, life-threatening infections such as HIV and hepatitis B, the file holder, long file holder,
and contrary electrode must be autoclaved after each patient’s treatment has been completed.
• Do not sterilize the autoclavable components by any method other than autoclaving.
• Components are extremely hot right after autoclaving. Wait for them to cool off before touching.
• Do not leave the components in the autoclave.