24
Instructions for Use 2019-03-20
Autoclavable Components
Wipe the file holder, long file holder and contrary electrode with
a piece of gauze dampened with Ethanol for Disinfection (Ethanol
70 vol% to 80 vol%).
Never wipe the contra angle with any solution other than Ethanol
for Disinfection (70 vol% to 80 vol%).
Do not immerse the components in or wipe it with any of the fol-
lowing: functional water (acidic electrolyzed water, strong alkaline
solution, and ozone water), medical agents (glutaral, etc.), medici-
nal solutions (FC: formalin cresol, sodium hypochloriteor, etc.) or
any other special types of water or commercial cleaning liquids.
Such liquids may result in plastic degradation, metal corrosion and
adhesion of the residual medical agent to the components. If any
of these liquids being applied to the components, wash it off in
running water.
• Do not use anything except ethanol (70 vol% to 80 vol%).
• Do not immerse the components in or wipe it with any of the following: functional water (acidic electrolyzed water, strong
alkaline solution, or ozone water), medical agents (glutaral, etc.), or any other special types of water or commercial clean-
ing liquids. Such liquids may result in metal corrosion and adhesion of the residual medical agent to the components.
• Never clean the components with chemicals such as formalin cresol (FC) and sodium hypochlorite. These will damage the
plastic parts of the components. If any of these liquids being applied to the components, wash it off in running water.
• Use only ethanol (70 vol% to 80 vol%) and OPTI-CIDE-3™ Surface Wipes for cleaning. Any other cleaning chemical or
products should not be used including but not limited to the following cleaning products and similar cleaning products
listed below because of the potential damage to the plastic components of the Endostar Navigator.
• CaviWipes™ • CaviCide™ • SANI-CLOTH™
Disinfection
Individually place the file holder or long file holder, and contrary electrode in a sterilization pouch.
Do not put stress on the cable when you place the file holder in a sterilization pouch.
Packing