44
Test Procedure
1. Press Shutter Button fully and note reading on the Stops Error Meter.
2. Repeat the procedure two more times, noting the Stops Error Meter readings.
Compare the readings to the Specification
3. If the test readings are not within Spec, adjust IR Strobe Exposure Calibration Slide
and retest.
4. If adjusting the IR Calibration Slide fails to bring the Strobe Exposure Test reading
within spec verify the following:
•
The IR black filter is seated
•
The photocell is seated in the base block
•
The photocell cap is seated properly.
•
If the above steps do not bring the Strobe Exposure (Graywall) reading within
spec, proceed to the LS Graywall Test.