POL-EKO RT 2014 Скачать руководство пользователя страница 22

Instruction manual RT 2014 

 

 

 

POL-EKO-

APARATURA sp.j. • ul. Kokoszycka 172 C • 44-300 Wodzisław Śl. • tel./fax +48/ 32 453 91 70 • e-mail 

[email protected]

 

22 

TECHNICAL DATA 

Parametr 

RT2014 

Temperature measurements 

2 x external Pt100 

Temperature range (depending on sensor

) [°C] 

-110 ... 400 (depending on type of sensor) 

Temperature accuracy 

[°C] 

0,1 / +/- 0,2 

Humidity measurement 

zewnętrzny RH_STD/RH_PREM 

Humidity range [%] 

RH_STD 0...80 RH_PREM 0...100 

Humidity accuracy [%] 

1% / RH_STD: 1,8; RH_PREM 0,8 

Atmospherical pressure measurement 

internal 

Pressure range [hPa] 

300 - 1100 

Pressure accuracy [hPa] 

1 / 1,5 

Sensor cable length [m] 

2,5 

Current time 

yes 

interval [min] 

1/5/15/30/60 

Internal memory 

Max 50 000 measurement data 

Additional memory 

microSD card up to 8GB 

interface 

Micro USB 

Power  

5 VDC ; 1A 

Ambient conditions 

10°C to 40°C 

up to 70 % RH 

Battery working time  

48 h 

display 

OLED 128x64 px 

External dimensions (w x d x h) [mm] 

72 x 20 x 85 (140 with antenna)  

weight [g] 

165 

Frequency band GSM 

850/900/1800/1900 

Quantity of phone numbers for sending SMS notification 

(ringing) 

warranty 

24 miesiące 

* in the table are the parameters for the most extended version of the model 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание RT 2014

Страница 1: ...RUCTION MANUAL Temperature data logger with text message notifications RT 2014 Caution Before using the device first read carefully this manual Manufacturer POL EKO APARATURA sp j Ver 1 49 Data 05 10...

Страница 2: ......

Страница 3: ...9 5 2 2 Installation of microSD card 10 5 2 3 Installation of GSM or WiFi antenna 10 5 2 4 Start up of the unit 10 5 2 5 Roaming 11 6 OPERATION OF THE DEVICE 11 6 1 Display 12 6 2 Actualization of th...

Страница 4: ...0 REMOTELY REQUESTED STATUS OF THE DEVICE 17 11 Accessories 18 11 1 Measurement sensors 18 11 2 Fitting 18 12 CLEANING AND MAINTENANCE 19 12 1 Handling with battery 20 13 RATING PLATE 21 TECHNICAL DAT...

Страница 5: ...tect the power cable and the plug from any damage and do not use the plug if it is improperly plugged in or if the cable is laid incorrectly disconnect the unit and protect it from reconnecting if it...

Страница 6: ...tional 2 x binary input output These notifications and functions relate to the most extended version of the RT2014 All devices are equipped with batteries capable to operate after a power failure Stan...

Страница 7: ...top cover In this purpose you have to 1 Unscrew 2 black screws visible on the photo 2 Take off the cover the button is located under microUSB slot 3 Proceed according to part 6 3 Reset of the device 4...

Страница 8: ...lative humidity max 70 applies to PT100 Handling of sensors sensors wires should be distanced from GSM or WiFi antenna at least 20cm sensors wires cannot be bended or coiled in the loop temperature se...

Страница 9: ...tion The first point is to activate the SIM card To activate the SIM card you should supply a mobile phone and using it you should 1 turn off the mobile phone 2 insert the SIM card in and turn on the...

Страница 10: ...ntenna WiFi Antenna 5 2 4 Start up of the unit After switching on the unit switching on is made by clicking and holding button GSM module will try to connect to mobile operator icon informs about look...

Страница 11: ...too weak to reliably operation In this case we it s recommended that you enable roaming services in GSM network The device does not distinguish domestic and abroad roaming For boarder areas the manuf...

Страница 12: ...nalling of alarm event the device is connected to a power source battery status growing symbols indicates battery charging flashing icon indicates lack of microSD card micro SD card icon continuous li...

Страница 13: ...in the following order 1 Plug micro USB cable to the charger 2 Plug the charger to the mains 3 If the device is completely discharged leave it to charg for 60min it s forbidden to push any button dur...

Страница 14: ...off the recorder Using the application AVIA The data on a memory card is stored in a folder which corresponds to the name of the data logger There are the following files plk format supported by suppl...

Страница 15: ...over the acceptable operating range of sensor Values from sensor 1 or 2 transcend over the acceptable operating range set in parameters alarms event and the time set in column delay passed by Return o...

Страница 16: ...the over temperature alarm specified in the column Low for the primary sensor RT2014 2014 04 18 12 55 End of ALARM low limit T 1 10 5 0C The alarm is signaled also by the device itself RT2014 in the f...

Страница 17: ...these conditions are fulfilled and the text message still will not be received it could mean a failure of measurement chain or data logger controller failure It is recommended to check the data logger...

Страница 18: ...40 180 C 2 5 PT100 L Sensor for low tem peratures recom mended for ZL temp 110 120 C 2 5 RH_STD Humidity and tempera ture sensor recom mended for ST IL RH 0 80 temp 0 60 C 2 5 RH_PREM Humidity and tem...

Страница 19: ...glue should be performed on a dry surface Handle adjustment mechanism allows to tilt the data logger Tilt adjustments should be per formed by setting the recorder at the desired position and then tig...

Страница 20: ...y for at least 60 minutes 12 1 Handling with battery The device is equipped with a maintenance free lithium polymer battery providing return of operation after power failure Replacing the battery shou...

Страница 21: ...ting plate is located on the back side of the device Here is an example of rating plates A without bracket B with fitting bracket Where 1 Manufacturer 2 Type of unit 3 Serial number 4 Type of used sen...

Страница 22: ...Atmospherical pressure measurement internal Pressure range hPa 300 1100 Pressure accuracy hPa 1 1 5 Sensor cable length m 2 5 Current time yes interval min 1 5 15 30 60 Internal memory Max 50 000 meas...

Страница 23: ...ing with sensors Probably dam aged sensor or measurement chain Contact service to check and replace broken element Despite the alarm notifica tion text mes sage call or both has not been made No sourc...

Страница 24: ...the user buyer is extended for the time the unit has spent at the ser vice 5 Warranty covers data carriers delivered along with the equipment warranty does not cover da ta saved on any additional data...

Страница 25: ...hoose position CDC and MSC right click and go to Update driver software Mark the location where are the files atmel_devices_cdc the location is on delivered DVD in Driver folder Copy the files to the...

Страница 26: ...is made in two stages STAGE 1 Disable the verification of digital signatures for Win8 Variant 1 pic 1 Move the cursor to the bottom right corner of the desktop Go to the power options and click on the...

Страница 27: ...172 C 44 300 Wodzis aw l tel fax 48 32 453 91 70 e mail info pol eko com pl 27 Choose Advanced options Choose Restart options Choose Restart The PC will boot and show options choose and click F7 butt...

Страница 28: ...STAGE 2 Drivers installation Click Start right click on This computer and choose Manage From the list choose Device manager Now choose Other devices right click on USB Serial Port and choose Update d...

Страница 29: ...DVD folder Driver atmel_devices_cdc Copy them on local disc e g C Then choose localization Computer and go Next Choose from hard disc and choose localizations where drivers atmel_devices_cdc Click Ope...

Страница 30: ...the driver will be installed Then Finish the installation In order to verify the proper operation of the con troller open window Device Manager as earlier in the manual Unroll Ports COM and LPT and c...

Страница 31: ...Instruction manual RT 2014 POL EKO APARATURA sp j ul Kokoszycka 172 C 44 300 Wodzis aw l tel fax 48 32 453 91 70 e mail info pol eko com pl 31...

Страница 32: ...Instruction manual RT 2014 POL EKO APARATURA sp j ul Kokoszycka 172 C 44 300 Wodzis aw l tel fax 48 32 453 91 70 e mail info pol eko com pl 32...

Страница 33: ...er the sole responsibility of the manufacturer Wymieniony powy ej przedmiot niniejszej deklaracji jest zgodny z odno nymi wymaganiami unijnego prawodawstwa harmonizacyjnego The object of the declarati...

Страница 34: ...Instruction manual RT 2014 POL EKO APARATURA sp j ul Kokoszycka 172 C 44 300 Wodzis aw l tel fax 48 32 453 91 70 e mail info pol eko com pl 34 Notes...

Страница 35: ......

Страница 36: ...igerators laboratory freezers heating ovens cooled incubators drying ovens colony counters stationary samplers specialized devices for controlled sewage and waste waters taking We organize trainings s...

Отзывы: