9
5.
Troubleshooting
Error
Possible cause
Possible solution
General Ceiling Hoist
error.
Loose Battery terminal(s).
Connect terminal(s).
Empty Batteries.
Replace the Batteries.
Hand Control connector loose.
Connect connector.
Hand Control is damaged.
Replace the Hand Control.
Lift moves down but not
up.
Limit Switch has been activated
Check the Limit Switch
Hand Control is damaged.
Replace the Hand Control.
Batteries are empty
Charge the Batteries
Lift moves up but not
down.
Lifting Tape tension switch has been
activated.
Check the switch.
No tension on the Lifting Tape.
Put tension on the Lifting Tape.
Hand Control is damaged.
Replace the Hand Control.
Lift has difficulties moving. Damaged wheel.
Replace the Wheel cap(s).
Lift makes (too much)
noise during lifting.
Dry Gear
Lubricate the Gear.
Dry Motor axle.
Disassemble the Motor and lubricate
the axle (advice; Kroon Oil Lithep
Grease EP2).
Lift does not charge.
Dirt on the charging contacts of the
Hand Control.
Clean the contacts.
Charger defect.
Replace the charger.
Hand Control damaged.
Replace the Hand Control.
Batteries damaged.
Replace the Batteries
Lift lacks lifting power.
Batteries are empty.
Charge the Batteries
Batteries damaged.
Replace the Batteries
Motor damaged.
Replace the Motor.
The mass drops slowly
without using the Hand
Controls.
Electronic Brake of the Motor is
damaged.
Replace the Electronic Brake.
Batteries run out quickly.
Batteries are empty
Charge the Batteries
Batteries damaged.
Replace the Batteries
Hand Control contacts are dirty
Clean the charging contacts
Hand Control is damaged.
Replace the Hand Control.