Neurotechnology Research Systems
© Plexon Inc
- 45 -
6.10
Start Mode
When the Start Mode control is in Individual mode, each channel and can be started or stopped
using the individual Start and Stop controls in the row of controls dedicated to that channel. Refer
to section 6.2 for a description of Stop and Start.
Pressing Stop All is a convenient way to stop the stimulation in progress on every channel. Pressing
Start All is a convenient way to start the stimulation protocol on every channel. Pressing Start All is
also the only way to start stimulation on multiple channels simultaneously from the GUI. Complex
multichannel stimulation protocols can be initiated using the Start All function. First define and
load the appropriate stimulation pattern for the channels you wish to stimulate and then initiate
the multichannel stimulation pattern by pressing the Start All button.
Note the pressing Stop All or Start All causes the individual channel Stop and Start controls to be
grayed out. Press the Individual button to start or stop channels independently.
Hint: You probably want to return the Start Mode to Individual after pressing Stop All or Start All
and there is no reason not to do so.
6.11
Monitor Channel
Use the dropdown menu to select one channel for display on the voltage and current monitor
outputs. Only one channel may be monitored a time. Connect the current and voltage monitor
outputs to an oscilloscope using standard BNC cables to see the stimulation pattern that is being
applied to the selected electrode.
Содержание Stimulator 2.0
Страница 59: ...Neurotechnology Research Systems Plexon Inc 59...