Step 9.
Remove the wash buffer lid from the bottle and place on clean paper towel.
Step 10.
Run the Rinse/Prime cycle of the Fluid Path on the setting menu four times. This step
will pump air through the tubing to displace any remaining liquid.
Step 11.
Remove the plastic U Tray (refer to section 6.1.1).
Step 12.
Run the Shutdown program to turn off the instrument.
Step 13.
Switch the power off.
Step 14.
Disconnect the power cord from the power plug.
Step 15.
Disconnect the Waste Bottle Lid and empty the Waste Bottle, rinse several times with
water.
Step 16.
Clean all lids with 70% ethanol and let air dry.
Step 17.
Clean all bottles with 70% ethanol and let air dry.
Step 18.
Assemble all bottle and lids with the IntelliPlex 1000 πCode Processor. Close the air
valves of the ddH
2
O and Wash Buffer lids.
7.2.5 Restarting after Long-term Shutdown
To initiate the IntelliPlex 1000 πCode Processor after Long-term Shutdown:
Step 1.
Open the air valves of the ddH
2
O and Wash Buffer lids.
Step 2.
Do not fill the ddH
2
O and Wash Buffer bottles.
Step 3.
Ensure that the on/off switch in the rear panel of the instrument is in the off position.
Step 4.
Plug the power cord into the back of the instrument and switch on the IntelliPlex
1000 πCode Processor.
Step 5.
Perform self-test (the system will pump air during self-test).
Step 6.
Place a new plastic U Tray (refer to section 6.1.1).
Step 7.
Fill the ddH
2
O and Wash Buffer bottles with at least 500mL ddH
2
O or Wash Buffer.
Step 8.
Perform Rinse/Prime two times (refer to section 6.4).
Visually confirm the uptake of ddH
2
O or Wash Buffer and confirm discharge into waste
bottle. If no ddH
2
O or Wash Buffer uptake is observed, repeat Rinse/Prime up to four
additional times. If still no uptake is observed, please contact Customer Service of
PlexBio Co., Ltd. or qualified local distributors. If no ddH
2
O or Wash Buffer uptake is
observed but no discharge into waste bottle, perform system shutdown, switch the
instrument and contact Customer Service of PlexBio Co., Ltd. or qualified local
distributors.
Step 9.
The IntelliPlex 1000 πCode Processor is operational.