45
Serial command:
R
Description: Recall conditions from memory.
Usage:
R(0-9)<cr>
Example:
Send “R2<cr>”. The beam conditions stored in memory location 2 will be recalled.
Acceptable range is from 0 to 9. Each memory location stores beam current, beam
voltage, accelerator voltage, discharge voltage, cathode filament current and
neutralizer filament current. Also stored are the A/B and E/B ratios, cathode and
neutralizer filament current limits. Memory location 0 is reserved for the last
running condition.
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
RC
Description: Request running conditions.
Usage:
RC<cr>
Example:
Send “RC<cr>”. The current running conditions will be sent to the buffer. It is
recommended this command should not be sent more than 1 time per second.
Response: <lf><cr>AA.A, B.BB,CCC.C,DDD,EEEE,FFF,GGGG,HHHH,II.I,JJJJ<lf><cr>
<eot>
Where: AA.A is cathode filament current (A)
B.BB is the discharge current (A)
CCC .C is the discharge voltage (V)
DDD is the beam current (mA)
EEEE is the beam voltage (V)
FFF is the accelerator current (mA)
GGGG is the accelerator voltage (V)
HHHH is the emission current (mA)
II.I is the neutralizer filament current (A)
JJJJ is factory use