25
The options for each module in the main display will utilize abbreviations. The following is a listing
of these abbreviations, the corresponding parameter and description. The specific module and
acceptable range are also provided. Consult the source check-out sheet for typical values.
MODULE ABBREVIATION PARAMETER AND DESCRIPTION
ACCEPTABLE
RANGE
Beam
BEAM V or
BV
Beam voltage
After the beam is turned on, the power
supply will ramp to the beam voltage.
0 to 1500 V
Beam
BI
Beam current
After the beam is turned on, the power
supply will ramp the beam current.
0 to 300 mA
Accelerator ACCEL
Accelerator voltage
Below 100 V is not recommended.
0 to 1000 V
Discharge
DISC V
Discharge voltage
Typical values are 35 to 55 V.
0 to 80 V
Cathode
CATH I
Cathode filament current
0.0 to 20.0 A
Neutralizer NEUT I
Neutralizer filament current
0.0 to 20.0 A
Control
SETUP
Select to change global parameters listed
below
Control
CATH I LIMIT
Cathode filament current limit
The limit should be set slightly higher than
the maximum expected cathode filament
current.
0.0 to 20.0 A
Control
NEUT I LIMIT
Neutralizer filament current limit
The limit should be set slightly higher than
the maximum expected neutralizer filament
current
0.0 to 20.0 A
Control
A/B RATIO
Accelerator to beam current ratio
If the accelerator current exceeds the A/B
ratio times the beam current, an alarm is
triggered. Default is 10%.
0 to 99%
Control
E/B RATIO
Emission to beam current ratio
Neutralizer emission will ramp to the E/B
ratio times beam current. Recommended is
125%.
0 to 200%
Control
RMT S/B SW
Remote source / beam switch
This will activate the rear panel remote
switch (see Chapter 5).
enable/disable
Control
ALARM
Audio alarm
enable/disable
Control
SECURITY
Factory use
MEM key
STORE
Store beam parameters into memory
0 to 9
MEM key
RECALL
Recall beam parameters from memory
0 to 9
All
INC
Increase value by 1 or 0.1
All
DEC
Decrease value by 1 or 0.1