7 INSTALLING THE X-RAY UNIT
26 Planmeca ProOne
Installation Manual
6. Connect the wires of the power cable as shown below.
7. Attach the strain relief to the junction box.
8. Attach the cover plate to the junction box.
NOTE
If you need to connect an existing power cable directly to the X-ray unit, refer to the wiring
diagram for an existing permanently installed power cable in section 11 "DIAGRAMS" on
PE
N
L
N
L
Br
o
w
n
Blue
Br
o
w
n
Blue
Y
e
llo
w
/ Gr
een
(ear
th)
Junction
box
Y
ello
w
/ Gr
een
(ear
th)
X-ray unit power cable
Mains
Junction box