Planmeca Proline EC Panoramic X-ray 41
TROUBLESHOOTING
User's manual
Er 07
The rotating unit is not in the ready position. Check that the program selection is com-
pleted. Press the ready key to move the rotating unit to the ready position and then
take an exposure.
Er 10
Overvoltage in tube head. Turn off the unit first for 30 seconds, try again without the
patient and the film cassette: take three exposures with values 60 kV and 4 mA and
one exposure with values 80 kV and 12 mA. If the error reappears after this, please
contact your service technician.
Er 11
Sudden kilovolt drop. Turn off the unit first for 30 seconds, try again without the patient
and the film cassette: take three exposures with values 60 kV and 4 mA and one expo-
sure with values 80 kV and 12 mA. If the error reappears after this, please contact
your service technician.
Er 12
Tube filament initializing is not done. Turn off the unit first for 30 seconds, try again
without the patient and the film cassette. If the error is still present, please contact your
service technician.
Er 21
Up/down motor time out. Check that the up/down movement is not obstructed. If there
is no visible obstruction, turn off the unit first for 30 seconds, try again without the
patient. If the error is still present, please contact your service technician.
Er 22
Temple rest motor time out. If this motor runs more than 3 seconds continuously this
error code occurs. Turn off the unit first for 30 seconds and then try again. If the error
is still present, the temple rest motor must be inactivated. Press the Control key and
press and hold down the Temple support key. The text HLd 0 will appear on the Main
display, and the supports can not be moved. The unit can be used without temple sup-
ports. Remove the supports and call your service technician.
Er 23
Layer adjust motor time out. Turn off the unit first for 30 seconds, try again without the
patient. If the error is still present, please contact your service technician.
Er 24
Primary slot motor time out. Turn off the unit first for 30 seconds, try again without the
patient. If the error is still present, please contact your service technician.
Er 25
Cassette motor time out. Try to drive the cassette carriage without the patient by
pressing the ready and return keys. Check that the movement is not obstructed. If
there is no visible obstruction, turn off the unit first for 30 seconds, try again. If the
error is still present, please contact your service technician.
Er 26
Rotation motor time out. Check that the rotation movement is not obstructed. If there is
no visible obstruction, turn off the unit first for 30 seconds, try again without the patient
and the film cassette. If the error is still present, please contact your service techni-
cian.
Er 30
kV value does not reach given value. Turn off the unit first for 30 seconds, try without
the patient and the film cassette. Take an exposure with values 60 kV and 4 mA and
an exposure with values 80 kV and 12 mA. If the error is after this present, please con-
tact your service technician.
Er 31
mA value does not reach given value. Turn off the unit first for 30 seconds, try without
the patient and the film cassette. Take an exposure with values 60 kV and 4 mA and
an exposure with values 80 kV and 12 mA. If the error is after this present, please con-
tact your service technician.
Er 32
Tube filament control inoperative. Turn off the unit first for 30 seconds, try without the
patient and the film cassette. Take an exposure with values 60 kV and 4 mA and an
exposure with values 80 kV and 12 mA. If the error is after this present, please contact
your service technician.
ERROR CODE
ERROR MESSAGE EXPLANATION