CLEANING
44 Planmeca Proline EC Panoramic X-ray
User's manual
14.3Cooling time
If a large number of exposures are taken in rapid suc-
cession the X-ray tube may overheat and stop work-
ing. A cooling time will appear on the Main display
which indicates the delay before next exposure can be
taken.
If the temperature of the tube head exceeds 60° C the
temperature will appear flashing on the display. The
unit will resume operation when the temperature drops
to the normal level. The time will then reappear on the
Main display.
15 CLEANING
NOTE
When disinfecting the unit surfaces, always
disconnect the unit from mains.
The bite piece and the patient support handles can be
cleaned with alcohol-based solutions. The bite piece
and patient support handles must be cleaned after
every patient.
Other unit surfaces can be cleaned with a soft cloth
damped in a mild cleaning solution. It is recommended
to clean the surfaces once a day.
16 SERVICE
To guarantee user and patient safety and to ensure
image quality the unit must be checked and recali-
brated by a qualified PLANMECA service technician
once a year or after every 10,000 exposures if this is
sooner. Please refer to the Planmeca Proline EC Pan-
oramic X-ray technical manual for complete servicing
information.