Mill
Symptom
Probable Cause
Corrective Action
Mill does not turn ON
Power cable unplugged
Power supply switch turned OFF
Faulty ON/OFF switch
Verify power cable is plugged into a live AC outlet.
Set power supply rocker switch at rear of mill to ON position.
Call Customer Support.
Block orientation on mill
incorrect.
Dirty Block Orientation Sensor
Using a small/long object, push very gently on the sensor switch
two or three times.
Pump making fast
thumping noise.
Dirty tank and/or filter
Clean tank and filter.
Air failure
message
Loss of air pressure
Air line loose
Ensure there is power to the office compressor.
Secure air line from compressor to mill.
Fluid Level Low Message
Low water supply
Faulty sensors
Fill tank with water.
Clean or replace sensors (Call Customer Support)
Tool stuck in collet
Collet rusty or caked with ceramics
Tool jammed due to air pressure
Remove tools manually. Done in the maintenance/command
screen on mill. (Call Customer Support for assistance).
Verify the supply air pressure 50 psig. Verify the pressure multiplier
is maxed out. i.e. the adjustment knob is turned fully clockwise
this doubles the pressure to 100 psig.
Not generating tool path Incomplete information from the Job Server
Restart Mill.
Milling is not completing
on the facial surface of an
anterior tooth
Restoration is longer than normal
On the design software, change the block selection to Multi Block,
but be sure to select the same block size. You can then insert a
standard Empress LT or HT block of the same size. The Mill makes
multiple passes on Multi blocks while other materials use a single
pass.
Unpleasant odor
Improper cleaning of the mil. Normal weekly
changing of the milling unit’s circulation
system coolant should be all that’s required to
keep the coolant tank clean and effective
A bleach flush can effectively correct the problem. Have bleach
on hand and ready, then call Customer Support for help with the
process.