10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
44 Planmeca ProMax 2D & 3D s & 3D Classic
User’s Manual (2D)
10.2.6 Selecting imaging angle
(Planmeca ProMax 2D S3 & Planmeca ProMax 3D units)
Use the arrow buttons in the middle of the screen to select
the imaging angle for this exposure.
NOTE
To change the default angle (17 degrees), select Settings
> Program > 2200 Program Features > Lateral TMJ
Default Angle or PA TMJ Default Angle and adjust the
setting with the plus or minus sign.
10.3 Patient positioning
10.3.1 Selecting patient entry position
Select the patient entry position as described in section
9.3.1 “Selecting patient entry position” on page 32.
10.3.2 Positioning patient in X-ray unit
1. Press the temple support button to open the temple
supports if they are not already open.
2. Guide the patient to the X-ray unit.
3. Adjust the X-ray unit to suit the height of the patient.
To do this, press either of the height adjusting
buttons until the opening in the chin support is
approximately level with the patient’s mouth.
+
-
Select imaging angle
Down
Up
Temple
support
button