10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
42 Planmeca ProMax 2D & 3D s & 3D Classic
User’s Manual (2D)
10.2.1 Selecting program
Select the TMJ program.
10.2.2 Selecting patient size
Select the patient size as described in section 9.2.2
“Selecting patient size” on page 28.
10.2.3 Adjusting exposure values for current exposure
The exposure values have been preset at the factory for
each patient size and program type. The preset exposure
values are average values and they are only meant to
guide the user.
NOTE
Always try to minimize the radiation dose to the patient.
The preset exposure values are shown in the following
tables.
Factory presets for lateral and lateral-PA TMJ exposures
PATIENT SIZE
kV VALUE
mA VALUE
Child (XS)
62
4
Small adult (S)
64
5
Medium-sized adult (M)
66
6.3
Large adult (L)
68
8
Extra large adult (XL)
70
10
Factory presets for posteroanterior TMJ exposures
PATIENT SIZE
kV VALUE
mA VALUE
Child (XS)
64
4
Small adult (S)
66
5
Medium-sized adult (M)
68
6.3
Large adult (L)
70
8
Extra large adult (XL)
72
10