6 MAIN PARTS
User’s Manual (2D)
Planmeca ProMax 2D & 3D s & 3D Classic 17
6.9 Patient positioning controls
NOTE
Never allow patients to press the positioning controls when
they are positioned in the X-ray unit.
NOTE
Pressing any of the positioning controls (button or joystick)
will switch the patient positioning lights on. The lights will
automatically switch off after two minutes. To switch them
off earlier, press the positioning joystick.
X-ray unit up / down
The X-ray unit up and down buttons are used to adjust the
X-ray unit to suit the height of the patient.
The X-ray unit moves slowly at first, then faster.
NOTE
If for some reason either of the buttons gets stuck during
operation, you can stop the up / down movement by
pressing any of the other control buttons or the positioning
joystick. This is a safety measure that guarantees that the
up / down movement can be stopped in an emergency.
NOTE
Be careful that the X-ray unit does not hit the ceiling when
you press the up button. The maximum height can be
adjusted to suit offices with low ceiling, contact your
service technician for help.
NOTE
Make sure that there is no object under the telescopic
column when you press the down button. If something is in
danger of becoming trapped, release the button
immediately to stop the movement.
Open / close temple supports
Move X-ray unit up
Positioning joystick
Move X-ray unit down
(2D imaging)
Down
Up