Figure 10: Alarm Settings window. 

6

 

Temperature Screening Location 

Preferred environmental conditions for the ThermaCheck M1 operation are 60 - 90˚F (16 - 32˚C) and less than 85% 

relative humidity.  The best screening location will have a mostly dark indigo color in the background of the 

thermal image.  If the background has a strong heat signature source, consider facing the ThermaCheck M1 camera 

in another direction.  Figure 11 illustrates some adverse environmental conditions that should be avoided in the 

screening area. 

 

Figure 11: Adverse environmental conditions for the ThermaCheck M1 camera 

Notes:   

1)

 

Temperature measurements of subjects entering the screening area from the outside could be affected if 

there is a significant temperature difference between these two environments. 

2)

 

When the environmental temperature in which the ThermaCheck M1 camera is operating changes by 

more than ±10 ˚F (±6 ˚C), the camera should be recalibrated for the new conditions.  If the user needs to 

operate the ThermaCheck M1 camera without a stable environmental temperature, it is recommended to 

use a blackbody in the scene of the camera to provide dynamic calibration. 

 

Avoid locations near vents,

 

exterior doors, and 

strong backlighting

 

For the best environmental stability 

 

pick a climate-controlled space

 

Содержание TC160M1

Страница 1: ...rface Functions 5 5 1 System settings and information 5 5 2 Interactive calibration function 5 5 3 Mask Detection 7 5 4 Screening Capture History 7 5 5 Temperature Threshold Setting 7 6 Temperature Sc...

Страница 2: ...subjects face The M1 Series camera may be used to screen subjects for wearing a mask issuing an alert in the case of noncompliance Recent individual screening results are summarized in a bottom pictur...

Страница 3: ...nector on the camera s rear pig tail 3 Connect power adapter to the wall outlet 4 Connect camera head to a display not provided using the HDMI cable provided 5 Connect provided mouse to camera head th...

Страница 4: ...4 ThermaCheck M1 main display Table 1 Main display features UI Feature Description 1 Cameras Realtime view of visible and thermal cameras 2 Screening Bar Dynamically refreshed recent screening results...

Страница 5: ...the mouse the appropriate display icon The ThermaCheck M1 camera settings are described below 5 1 System settings and information System Tab 1 Selection of the date time field opens a dialog box for...

Страница 6: ...A approved device is preferred If using a blackbody subject set setpoint temperature to 95 97 F 35 36 C and allow the blackbody to reach the setpoint temperature before calibration is started 3 Use th...

Страница 7: ...on choose the date and screening record type Normal High Temp No Mask to perform a search of the screening record as shown in Figure 9 Figure 9 Selecting record type for search of the screening record...

Страница 8: ...rea Figure 11 Adverse environmental conditions for the ThermaCheck M1 camera Notes 1 Temperature measurements of subjects entering the screening area from the outside could be affected if there is a s...

Страница 9: ...le temperature source used in the factory calibration of thermal cameras to make their temperature measurements as accurate as possible Some infrared cameras used for human temperature screening requi...

Страница 10: ...rature is influenced by environmental conditions Individuals with abnormal skin temperature readings should be further evaluated with a medical grade thermometer Planck Vision Systems is not advertisi...

Отзывы: