routines that give ThermaCheck M1 superior temperature stability and measurement accuracy. We recommend 

that our cameras be calibrated on the human face, using a medical grade thermometer. 

 

Figure 13: Stability of ThermaCheck M1 camera shown over a 30hr period. 

Note: 

When the environmental temperature in which the ThermaCheck M1 camera is operating changes by 

more than ±10 ˚F (±6 ˚C), the camera should be recalibrated for the new conditions.  If the user needs to 

operate the ThermaCheck M1 camera without a stable environmental temperature, it is recommended to 

use a blackbody in the scene of the camera to provide dynamic calibration. 

9

 

Can ThermaCheck M1 be used to diagnose illness? 

No. The ThermaCheck M1 camera is designed for preliminary temperature screening only. Although the available 

scientific literature supports the use of infrared cameras for this purpose [1], they cannot be used for medical 

diagnostics since the skin temperature is influenced by environmental conditions. Individuals with abnormal skin 

temperature readings should be further evaluated with a medical grade thermometer. 

Planck Vision Systems is not advertising our cameras as medical equipment. Our products can only identify 

individuals with elevated skin temperature. There is no way to thermally detect an infected individual who does 

not have an elevated body or skin temperature and only a licensed medical professional can determine if such an 

individual is experiencing an abnormal medical condition. 

_____________________ 

[1] Ring, Francis J., and E. Y. K. Ng. “Infrared thermal imaging standards for human fever detection.” 

Medical 

Infrared Imaging: Principles and Practices

. CRC press, 2007.   

10

 

Thank you 

Thank you for purchasing a ThermaCheck M1 camera.  We develop our products to deliver reliable, long lasting 

performance, and want to make sure you are satisfied with your purchase.  For more product information visit 

www.planckvisionsystems.com

, and feel free to get in touch with us at 

[email protected]

 with 

questions, problems, and comments. 

 

5

 

10

 

98.

8

 

97.

 

T (˚F)

 

± 0.5 ˚F

 

0

 

1

5

 

3

0

 

2

0

 

2

5

 

Actual Temperature (98.1 ± 0.2 ˚F)

 

Therma

Check

TM

 

Time (hr)

 

Содержание TC160M1

Страница 1: ...rface Functions 5 5 1 System settings and information 5 5 2 Interactive calibration function 5 5 3 Mask Detection 7 5 4 Screening Capture History 7 5 5 Temperature Threshold Setting 7 6 Temperature Sc...

Страница 2: ...subjects face The M1 Series camera may be used to screen subjects for wearing a mask issuing an alert in the case of noncompliance Recent individual screening results are summarized in a bottom pictur...

Страница 3: ...nector on the camera s rear pig tail 3 Connect power adapter to the wall outlet 4 Connect camera head to a display not provided using the HDMI cable provided 5 Connect provided mouse to camera head th...

Страница 4: ...4 ThermaCheck M1 main display Table 1 Main display features UI Feature Description 1 Cameras Realtime view of visible and thermal cameras 2 Screening Bar Dynamically refreshed recent screening results...

Страница 5: ...the mouse the appropriate display icon The ThermaCheck M1 camera settings are described below 5 1 System settings and information System Tab 1 Selection of the date time field opens a dialog box for...

Страница 6: ...A approved device is preferred If using a blackbody subject set setpoint temperature to 95 97 F 35 36 C and allow the blackbody to reach the setpoint temperature before calibration is started 3 Use th...

Страница 7: ...on choose the date and screening record type Normal High Temp No Mask to perform a search of the screening record as shown in Figure 9 Figure 9 Selecting record type for search of the screening record...

Страница 8: ...rea Figure 11 Adverse environmental conditions for the ThermaCheck M1 camera Notes 1 Temperature measurements of subjects entering the screening area from the outside could be affected if there is a s...

Страница 9: ...le temperature source used in the factory calibration of thermal cameras to make their temperature measurements as accurate as possible Some infrared cameras used for human temperature screening requi...

Страница 10: ...rature is influenced by environmental conditions Individuals with abnormal skin temperature readings should be further evaluated with a medical grade thermometer Planck Vision Systems is not advertisi...

Отзывы: