Planck VISION SYSTEMS TC160M1 Скачать руководство пользователя страница 5

 

 

 

TC160M1  

 

User Manual

 

(v1.3)

 

www.planckvisionsystems.com 

 

5

 

User Interface Functions 

Screening parameters for the ThermaCheck M1 camera may be set through the user interface when a mouse is 

connected to the USB port provided in the camera’s rear pig-tail.  Access to settings are obtained by selecting, with 

the mouse, the appropriate display icon.  The ThermaCheck M1 camera settings are described below. 

5.1

 

 System settings and information 

System Tab

:  

1)

 

Selection of the date/time field opens a dialog box for setting this field. 

2)

 

Desired temperature unit is selected with the corresponding radio button. 

 

  

 

Figure 5: System settings window 

Version Parameter Tab

:  

Opens widow providing system information. 

 

 

 

Figure 6: System version window 

5.2

 

 Interactive calibration function 

Starts an interactive calibration procedure.  

Содержание TC160M1

Страница 1: ...rface Functions 5 5 1 System settings and information 5 5 2 Interactive calibration function 5 5 3 Mask Detection 7 5 4 Screening Capture History 7 5 5 Temperature Threshold Setting 7 6 Temperature Sc...

Страница 2: ...subjects face The M1 Series camera may be used to screen subjects for wearing a mask issuing an alert in the case of noncompliance Recent individual screening results are summarized in a bottom pictur...

Страница 3: ...nector on the camera s rear pig tail 3 Connect power adapter to the wall outlet 4 Connect camera head to a display not provided using the HDMI cable provided 5 Connect provided mouse to camera head th...

Страница 4: ...4 ThermaCheck M1 main display Table 1 Main display features UI Feature Description 1 Cameras Realtime view of visible and thermal cameras 2 Screening Bar Dynamically refreshed recent screening results...

Страница 5: ...the mouse the appropriate display icon The ThermaCheck M1 camera settings are described below 5 1 System settings and information System Tab 1 Selection of the date time field opens a dialog box for...

Страница 6: ...A approved device is preferred If using a blackbody subject set setpoint temperature to 95 97 F 35 36 C and allow the blackbody to reach the setpoint temperature before calibration is started 3 Use th...

Страница 7: ...on choose the date and screening record type Normal High Temp No Mask to perform a search of the screening record as shown in Figure 9 Figure 9 Selecting record type for search of the screening record...

Страница 8: ...rea Figure 11 Adverse environmental conditions for the ThermaCheck M1 camera Notes 1 Temperature measurements of subjects entering the screening area from the outside could be affected if there is a s...

Страница 9: ...le temperature source used in the factory calibration of thermal cameras to make their temperature measurements as accurate as possible Some infrared cameras used for human temperature screening requi...

Страница 10: ...rature is influenced by environmental conditions Individuals with abnormal skin temperature readings should be further evaluated with a medical grade thermometer Planck Vision Systems is not advertisi...

Отзывы: