4.5.1.6.
4.5.1.7.
4.5.1.8.
Checking Removable Module
This program checks the condition of the Removable Module components. If the test
does not detect any error, the green indicator will be displayed in the status field. If an
error has been detected, the indicator will be red, and in this case the Meter should be
sent to repair service to have the defects corrected.
Removable Module Temperature
The parameter is intended for checking the temperature inside the Removable
Module. Temperature is the main environmental factor, which influences the accuracy of
measurements. To check the accuracy of measurements of the signal level, use the
temperature parameter. If the temperature value is within the allowed range (from – 10 to +
50
0
С
), then the green indicator will be displayed in the status field. If the temperature is
beyond this allowed range, then the indicator will be red.
Calibration
This program checks the condition of the correction factor tables. If the test does not
detect any error, the green indicator will be displayed in the status field. If an error has
been detected, the indicator will be red, and in this case the Meter should be sent to repair
service to have the defects corrected.
4.5.2. Setting Regional Parameters
Regional parameter setting is intended for the Meter configuration in accordance to
the local conditions and for the user convenience. The
icon in the main menu
corresponds to this mode. The screen view is shown in figure 4.23.
Figure 4.23
The table contains the following parameters which are available for editing:
1.
Language
. Selecting of the user interface language.
2.
Ch template
. Selecting of the channel allocation standard.
3.
Level units
. Selecting of the signal level measurement units.
4.
Time format
. Selecting of the date representation format.
5.
Date format
. Selecting of the time representation format.
6.
Temperature units
. Selecting of the temperature measurement units.
Use and keys to select the required parameter, and set its value by pressing
◄
and
►
keys.
4.5.3. User Interface Setting
The operating parameter configuration program is intended for modifying the general
parameters of the Meter operation. The
icon in the main menu corresponds to this
program. See the parameter configuration screen view in the figure 4.24:
25