11.3 Placing the system in standby mode
While fault finding it is useful to place the DPM Reader in standby mode where it will
stop taking samples. To do this navigate to the SYSTEM EVENT SETTINGS page and
change the bSystemSample value in the COUNT* column to 0 seconds. This will stop
the DPM Reader from sampling until the DPM Reader is powered off then on again or
the 0 is changed to 240 seconds to start the countdown for the next sample. Remember
to reactivate this once you have finished fault finding the DPM Reader.