Installation
Warning: Set DPM Reader to Safety Control Level 4
1. Secure heater sub-assembly with 2 new nyloc nuts.
2. Insert cable in to DPM carrier board heater header.
3. RTV connector on DPM carrier board.
9.2.6
05-0791 Air Flow Sensor Assembly Replacement
Removal
Warning: Set DPM Reader to Safety Control Level 4
Refer to Figure 42
F
IGURE
42
-
R
EMOVAL OF
A
IR
F
LOW
S
ENSOR
1. Cut cable tie.
2. Unplug connector.
3. Using No.1 Phillips head screwdriver, undo the 4 screws and washers.
Installation
Warning: Set DPM Reader to Safety Control Level 4
1. Secure air flow sensor with 4 screws.
2. Connect cable to Air flow sensor.
3. Secure cable tie onto sticky mount on top of air flow sensor.
4. RTV connector on Air Flow sensor
5. Complete flow sensor calibration.
9.2.7
Blower Speed Controller Replacement
Removal
Step 1
Step 3
Step 2