Operation
44
© 2017 Pico Technology
pv106ug r1
If two idential DUTs are available, a conventional isolation
calibration step can be carried out as shown
6.3
Measurements
6.3.1
Return loss
In order to carry out return loss measurements (S
11
) the VNA must be calibrated as described before, either
S
11
calibration only, both S
11
and S
21
or 12-term. The device to be tested (DUT) is then connected to Port 1 of
the VNA as indicated in the figure below.
If the DUT can only be connected to the VNA using a cable, then the
VNA should be calibrated at the end of the cable for best results.
Connect device under test to Port 1 to carry out S11
measurements.