5 Installation
22
Version: 1.2.0
PZ244E
P54x Nanopositioner
NOTICE
Center of load at unsuitable position!
If the center of load is located too far away from the center of the platform (e.g., tall loads and
unwanted lever effect), the P-54x can be damaged by high strain on the flexure guides, high
torques, and oscillation, especially when operated dynamically.
If the center of the load to be affixed is too far above or to the side of the platform, adjust
the controller settings before startup or contact our customer service department (p. 33).
NOTICE
Excessively long screws!
The P-54x could be damaged by screws inserted too deeply.
Pay attention to the depth of the mounting holes in the platform (p. 40).
Use screws of the correct length for the respective mounting holes only.
INFORMATION
The positive direction of motion is specified in the dimensional drawing (p. 40) by arrows.
Center of load at the optimal position:
Figure 5: Example of an optimally placed load
Center of load at an unsuitable position:
Figure 6: Tall load and center of load too far above the platform