2. Drive screw
Auxiliary Materials Required
■
Lubricant for the drive screw: Isoflex Topas 30
■
Lubricant for the guides: Klüberquiet BQ 72
‑
72
Relubricating
Under laboratory conditions, it is only necessary to relubricate the L-505.02A212F in
exceptional cases. For continuous industrial use, the lubrication intervals must be defined
individually.
If you have any questions on relubricating, contact our
customer service department (p. 30)
.
8.3
Moving the Motion Platform by Hand
It can be necessary to move the motion platform manually to provide access to mounting
holes for mounting screws in the base body of the positioner.
Requirements
✓
You have disconnected the L-505.02A212F from the electronics.
Moving the Motion Platform by Hand
1. Insert a TX 6 hex key into face of the drive screw.
2. Turn the hex key as far as necessary.
a) Turn the screwdriver clockwise to move the motion platform in the direction of the
screw hole.
b) Turn the screwdriver counterclockwise to move the motion platform away from the
screw hole.
➔
The rotary motion is transferred directly to the drive screw.
8.4
Cleaning
Requirements
✓
You have disconnected the L-505.02A212F from the electronics.
Auxiliary Materials Required
■
Soft, lint-free cloth
■
Mild cleaning agent or disinfectant
8 MAINTENANCE
L505M0020EN
‒
10/22/2018