You have mounted the L-417.xx9025E1
✓
The L-417.xx9025E1 is not connected to the electronics.
✓
You have prepared the load so that it can be fixed to the mounting holes in the motion
platform:
■
The gap between the center of gravity of the load and the center of the motion
platform is as small as possible in all directions.
■
At least three points are provided for fixing the load on the motion platform.
Mounting the Load onto the L-417.xx9025E1
1. If necessary: Insert the locating pins into the corresponding holes in the motion platform.
2. Align the load on the motion platform so that the mounting holes selected in the platform
can be used to mount the load.
3. Place the load onto the motion platform so that the locating pins are inserted into the
corresponding locating holes in the load.
4. Tighten the screws in all mounting holes.
5. Check that the load is sitting firmly on the motion platform.
6.4
Connecting the L-417.xx9025E1
NOTICE
Damage due to incorrect connection of the L-417.xx9025E1!
Connecting unsuitable electronics or the wrong cable can lead to damage to the
L-417.xx9025E1 or the electronics.
►
Make sure that the electronics support the drive type of the L-417.xx9025E1 and have been
configured accordingly.
►
Use cables from PI miCos only to connect the L-417.xx9025E1 to the electronics.
►
Pay attention to correct
Tools and Accessories
■
■
Cable set suitable for the electronics used
■
If necessary: Suitable screwdriver for the locking screws of the connectors.
■
Optional: Purge air connector with the following properties:
Pressure: 2 to 6 bar (200 to 600 kPa); purge air hose: outer diameter 4 mm; air quality as
stated in the
Requirements
✓
You have read and understood the
general safety instructions (p. 8)
✓
If necessary: The L-417.xx9025E1's base body is connected to the
.
✓
You have read and understood the user manual for the electronics.
✓
You have installed the electronics properly.
✓
The electronics are switched off.
Connecting the L-417.xx9025E1
1. Remove the ESD protective caps from all connectors on the L-417.xx9025E1.
2. All of the L-417.xx9025E1's connections must be made in the following order:
a) Optional: Purge air connector
Permissible pressure: 2 to 6 bar (200 to 600 kPa)
6 INSTALLATION
L417UM0002EN
‒
10/25/2022
21
M O T I O N | P O S I T I O N I N G