Technical Data
Ambient Conditions and Classifications
The following ambient conditions must be observed for the F-712 systems:
Area of application
For indoor use only
Maximum altitude
2000 m
Air pressure
1100 hPa to 0.1 hPa
(corresponds to roughly 825 Torr to 0.075 Torr)
Relative humidity
Highest relative humidity 80 % for temperatures up to 31 °C
Decreasing linearly to 50 % relative humidity at 40 °C
Operating temperature
Mechanics: –20 °C to 65 °C
Electronics: 5 °C to 40 °C
Storage temperature
0 °C to 65 °C
Transport temperature
–20 °C to 65 °C
Overvoltage category
II
Protection class
I
Degree of pollution
1
Degree of protection according to
IEC 60529
IP20
Maximum Ratings
The M-122K025 linear stages are designed for the following operating data:
Maximum
operating
voltage
Maximum
operating frequency
(unloaded)
Maximum power
consumption
12 V DC
2.25 W
The P-616K001 NanoCube® nanopositioner is designed for the following operating data:
Maximum
operating
voltage
Maximum
operating frequency
(unloaded)
Maximum power
consumption / axis
-20 to +120 V
240 Hz / X axis
240 Hz / Y axis
240 Hz / Z axis
4.3 W / X axis
4.3 W / Y axis
4.3 W / Z axis