6 Installation
C-863 Mercury Controller
MS205Equ
Version: 2.0.0
35
6.5 Connecting the Stage
NOTICE
Damage if a wrong motor is connected!
Connecting a stage with stepper motor to a DC motor controller can cause
irreparable damage.
Only connect to the C-863 a stage with DC motor or voice coil drive.
INFORMATION
The C-863 supports stages with a PWM amplifier as well as stages without a PWM
amplifier. Separate lines on the
DC Motor only
socket (p. 97) are available for both
stage versions. The selection of the proper lines must be ensured via the connector
of the stage. With stages from PI the proper connection is ensured.
Prerequisite
The C-863 is switched off, i. e. the power supply is
not
connected to the
power socket with the power cord.
Tools and accessories
If the distance between the C-863 and the stage is too great:
motor cable C-815.38, 3 m, available as an optional accessory (p. 16)
Connecting the stage
Connect the stage to the
DC Motor only
socket of the C-863.