PHYSITEMP TH-5 Скачать руководство пользователя страница 7

7.0   USE OF THERMALERT WITH A CHART RECORDER

7.1 

Analog output is optional on Model TH-5.

WARNING: Use of AC-powered equipment with Physitemp Thermometers. As it is powered
entirely by low-voltage batteries, Physitemp Thermometer TH-5 presents no safety hazard to
patients. However, caution is advised when using the TH-5 with a chart recorder for recording
patient temperature. Such auxiliary equipment, when AC-powered and in faulty condition, can
create a potential shock hazard.

Hazard to a Grounded Patient
The circuit of the TH-5 is such that a low-impedance path exists from the analog output socket
to the probe input socket. The TH-5 with sensor probe could therefore act as an electrical con-
ductor between a faulty AC-powered recorder and the patient. Leakage current could exceed
safe limits.

Hazard to an Ungrounded Patient
Some chart recorders have a permanently grounded input binding post. If a patient is connected
to faulty AC-powered equipment and becomes live, a leakage path to ground would exist
between the sensor probe and the recorder. Current safety limits could be exceeded.

If recording is essential:

Use a recorder which meets AAMI safe current limit standards. 
Before connecting thermometer to recorder, consult a competent 
engineer about shock hazard. Remove connector plug from ther-
mometer analog output when not in use.

7.2 

A DC output with amplitude approximately proportional to instrument reading is provid-

ed from a stereo-type jack on the back panel.  A mating plug is supplied for connection to the
recorder.  This plug may be left in the socket until it is required for use. The output of the TH-5
is 2 kilohms impedance, so connection to a potentiometric recorder will not affect either the
calibration or accuracy.

Connection diagram for the plug:

7.3  

Recorder sensitivity and pen zero must be adjusted so that the lines on the chart corre-

spond to the desired range of temperatures

7

NEGATIVE (RETURN)

POSITIVE (SIGNAL, OUTPUT)

Содержание TH-5

Страница 1: ...L INSTALLATION 3 0 PRECAUTIONS 4 0 DIRECTIONS FOR USE 5 0 REPLACING BATTERIES 6 0 CHECKING CALIBRATION 7 0 USE OF THERMALERT WITH CHART RECORDER 8 0 SPECIFICATIONS 9 0 TEMPERATURE MEASUREMENT WITH THE...

Страница 2: ......

Страница 3: ...rature monitoring system use suitable accepted technique during eletro surgical procedures to minimize the possibility that the probe cable will serve as an alternate path for radio frequency current...

Страница 4: ...e or probe Check connection If this is good replace probe A flashing colon in the center of the display signifies low battery alert Accurate readings will still be obtained and it is not necessary to...

Страница 5: ...3 4 full Add distilled water Leave for 30 minutes to allow temperature to stabilize stirring periodically b Immerse the standard mercury in glass thermometer in the ice bath to the immersion mark on t...

Страница 6: ...ent display and possible probe error readings should agree within the limits noted in Section 9 7 This means as an approximate rule of thumb that differences between standard and instrument readings s...

Страница 7: ...ded input binding post If a patient is connected to faulty AC powered equipment and becomes live a leakage path to ground would exist between the sensor probe and the recorder Current safety limits co...

Страница 8: ...y may be used Carbon Zinc cells are not recommended Battery Life Expectancy 1200 hours continuous alkaline Analog Output Optional extra approx 25mV C Size 3 H x 6 1 W x 6 1 4 D Weight 2 1 2 lbs Leakag...

Страница 9: ...raceable to NIST calibrations e Low cost users can even make their own sensors f Microscopic size when needed as in Physitemp microprobes g Nearly instant response h Better measurement accuracy due to...

Страница 10: ...This simple precaution must always be taken 9 9 Measurements of Air Temperature Temperature can vary widely in different parts of a room differences of at least several degrees will usually be noted W...

Страница 11: ...information by telephone at Tel 973 779 5577 Fax 973 779 5954 E mail physitemp aol com 10 3 Physitemp Instruments Inc warrants this instrument to be free from defects in material and workmanship for...

Отзывы: