4.0   DIRECTIONS FOR USE

4.1 

Connect any Physitemp type T thermocouple probe to the blue polarized socket on the rear 

of the instrument.

4.2  

Switch ON by depressing switch on front of unit - switch shows orange when in ON position.

For several seconds the digital display should read “1888” as a segment check. Do not use if any dis-
play segment is inoperative. A calibration check is not necessary before each use.

4.3  

Read temperature. Leave on for monitoring as long as desired.

4.4  

Switch off after use.

4.5  

Special Display Indications

1

A single digit 1 to the left signifies reading is outside
instrument range.

0

A single digit 0 to the left indicates an open cable or
probe.  Check connection.  If this is good, replace probe.

A flashing colon in the center of the display signifies

low battery alert.  Accurate readings will still be
obtained, and it is not necessary to discontinue use
immediately.  Install new batteries the next time the
unit is turned off.

A non-flashing colon signifies imminent battery

failure.  Install new batteries to ensure accurate
readings.

5.0  

REPLACING BATTERIES

5.1  

CAUTION: Use only alkaline “C" cell batteries. Switch instrument OFF.

5.2  

Grasp "squeeze release" on back panel and pull off battery retainer.

5.3  

Remove battery connectors and slide out battery holder.

5.4  

Discard old batteries and replace with fresh “C" cells.

5.5  

Replace battery holder and refit connectors.

5.6  

Insert top edge of battery retainer into lip of case. While grasping "squeeze release" push

retainer in until flush with back of case.

4

:

:

Содержание TH-5

Страница 1: ...L INSTALLATION 3 0 PRECAUTIONS 4 0 DIRECTIONS FOR USE 5 0 REPLACING BATTERIES 6 0 CHECKING CALIBRATION 7 0 USE OF THERMALERT WITH CHART RECORDER 8 0 SPECIFICATIONS 9 0 TEMPERATURE MEASUREMENT WITH THE...

Страница 2: ......

Страница 3: ...rature monitoring system use suitable accepted technique during eletro surgical procedures to minimize the possibility that the probe cable will serve as an alternate path for radio frequency current...

Страница 4: ...e or probe Check connection If this is good replace probe A flashing colon in the center of the display signifies low battery alert Accurate readings will still be obtained and it is not necessary to...

Страница 5: ...3 4 full Add distilled water Leave for 30 minutes to allow temperature to stabilize stirring periodically b Immerse the standard mercury in glass thermometer in the ice bath to the immersion mark on t...

Страница 6: ...ent display and possible probe error readings should agree within the limits noted in Section 9 7 This means as an approximate rule of thumb that differences between standard and instrument readings s...

Страница 7: ...ded input binding post If a patient is connected to faulty AC powered equipment and becomes live a leakage path to ground would exist between the sensor probe and the recorder Current safety limits co...

Страница 8: ...y may be used Carbon Zinc cells are not recommended Battery Life Expectancy 1200 hours continuous alkaline Analog Output Optional extra approx 25mV C Size 3 H x 6 1 W x 6 1 4 D Weight 2 1 2 lbs Leakag...

Страница 9: ...raceable to NIST calibrations e Low cost users can even make their own sensors f Microscopic size when needed as in Physitemp microprobes g Nearly instant response h Better measurement accuracy due to...

Страница 10: ...This simple precaution must always be taken 9 9 Measurements of Air Temperature Temperature can vary widely in different parts of a room differences of at least several degrees will usually be noted W...

Страница 11: ...information by telephone at Tel 973 779 5577 Fax 973 779 5954 E mail physitemp aol com 10 3 Physitemp Instruments Inc warrants this instrument to be free from defects in material and workmanship for...

Отзывы: