6.0   CHECKING CALIBRATION

6.1  

The Thermalert TH-5 thermometer incorporates an automatic zeroing amplifier which produces

near-perfect stability of calibration.  A calibration check is not necessary before each use.  Under nor-
mal circumstances, the instrument will provide years of service without requiring recalibration.

6.2 

Should it be suspected that the instrument is out of calibration, it may be checked with the

Physitemp type T calibrator (Model CT-3), or if this is unavailable, with an NIST calibrated thermome-
ter.  Verifying instrument calibration with an NIST calibrated thermometer requires extremely careful
technique, an accurate reference standard thermometer, and a reasonably draft-free location.  The fol-
lowing apparatus and supplies are needed:

1 - quart thermos flask with 2-hole cork
Stirrer
NIST traceable mercury in glass thermometer, 0.05°
graduations, partial immersion, 25-50°C (78-122°F)
with ice point scale.
Physitemp type T temperature probe
Ice made from distilled water
Distilled water

6.3  

Preparation:

a) Crush the ice and place it in the thermos flask, at least 3/4 full. Add distilled water. Leave for 
30 minutes to allow temperature to stabilize, stirring periodically.

b) Immerse the standard mercury-in-glass thermometer in the ice bath to the immersion 
mark on the stem. Stir the bath well and compare with thermometer reading at 0°C with the 
correction table supplied with the thermometer. If the error is greater than 0.0°C, wait 10 
minutes, stir the water bath and remeasure. If the error remains greater, test with another 
standard thermometer.

6.4  

A complete calibration check requires two different measurements, one for each of the

temperature ranges of the TH-5.

6.5  

Checking low end of temperature range:

Fill a thermos flask with a mixture of hot and cold water to produce a temperature of 
approximately 25

°

C.  Attach any Physitemp type T thermocouple probe to the standard 

mercury thermometer so that the sensor is at the same level as the mercury bulb, and 
immerse in the flask to the immersion depth marked on the stem. Do not allow probe or 
thermometer to touch the sides of the flask. Connect the probe to the TH-5 and switch it 
ON. Stir the bath well and wait 5 - 10 minutes. Ensure the flashing colon (battery warning) 
is not visible before reading the instrument and then the mercury thermometer to .05°C.
Make a note of both readings.

5

Содержание TH-5

Страница 1: ...L INSTALLATION 3 0 PRECAUTIONS 4 0 DIRECTIONS FOR USE 5 0 REPLACING BATTERIES 6 0 CHECKING CALIBRATION 7 0 USE OF THERMALERT WITH CHART RECORDER 8 0 SPECIFICATIONS 9 0 TEMPERATURE MEASUREMENT WITH THE...

Страница 2: ......

Страница 3: ...rature monitoring system use suitable accepted technique during eletro surgical procedures to minimize the possibility that the probe cable will serve as an alternate path for radio frequency current...

Страница 4: ...e or probe Check connection If this is good replace probe A flashing colon in the center of the display signifies low battery alert Accurate readings will still be obtained and it is not necessary to...

Страница 5: ...3 4 full Add distilled water Leave for 30 minutes to allow temperature to stabilize stirring periodically b Immerse the standard mercury in glass thermometer in the ice bath to the immersion mark on t...

Страница 6: ...ent display and possible probe error readings should agree within the limits noted in Section 9 7 This means as an approximate rule of thumb that differences between standard and instrument readings s...

Страница 7: ...ded input binding post If a patient is connected to faulty AC powered equipment and becomes live a leakage path to ground would exist between the sensor probe and the recorder Current safety limits co...

Страница 8: ...y may be used Carbon Zinc cells are not recommended Battery Life Expectancy 1200 hours continuous alkaline Analog Output Optional extra approx 25mV C Size 3 H x 6 1 W x 6 1 4 D Weight 2 1 2 lbs Leakag...

Страница 9: ...raceable to NIST calibrations e Low cost users can even make their own sensors f Microscopic size when needed as in Physitemp microprobes g Nearly instant response h Better measurement accuracy due to...

Страница 10: ...This simple precaution must always be taken 9 9 Measurements of Air Temperature Temperature can vary widely in different parts of a room differences of at least several degrees will usually be noted W...

Страница 11: ...information by telephone at Tel 973 779 5577 Fax 973 779 5954 E mail physitemp aol com 10 3 Physitemp Instruments Inc warrants this instrument to be free from defects in material and workmanship for...

Отзывы: