Notes on Operation
3
Notes on Operation
3.1
Connection and Start-up
Connecting the unit
(1)
Check whether the operating voltage of the device (see red window in the
mains module
)
<1>
) and the line voltage correspond.
(2)
Plug the supplied mains lead firmly into the rear panel of the device (
mains
module <1>
) and connect it to the socket. A safe connection is established via
a grounding socket outlet with ground wire.
3.2
Device start-up
Note
Prior to switching on the device, please make sure that the activated
ultrasonic transducer is outside the water in a clean condition.
(3)
Make sure that the instrument is switched off.
(4)
Turn the
dose control <6>
to 0.
(5)
Plug the transducers into the
transducer connector <11>
.
(6)
Switch on the device with the mains switch (
mains module <1>)
.
The instrument will carry out an automatic function check during which all
functions and output values are checked and the transducer is calibrated.
An audible signal is issued.
The instrument is ready now
. The parameters of the last treatment are preset.
3.3
Function check
(1)
Turn the
dose control <6>
to 0.
(2)
Switch the device off and on with the mains switch (
mains module <1>
).
The instrument will carry out an automatic function check during which all
functions and output values are checked and the transducer is calibrated.
An audible signal is issued.
The instrument is ready now
. The parameters of the last treatment are preset.
PHYSIOSON-Expert
9