Technical Data
13
Technical Data
Protection class
I acc. to VDE 0750 / IEC 601
BF type
CE characterization
acc. to Council Directive concerning medical devices
93/42 EEC
Class acc. to EC 93/42
IIa
Mains connection
230 VAC ± 10% or 115 VAC ± 10%
Observe the voltage selected at the mains module
<1>
Power line frequency
50-60 Hz
Power line input
66 VA
Ambient temperature
+ 10°C ... + 40°C
Line fuses
at 230 V:
T
500 mA
at 115 V:
T 1 A
Output data
small transducer
large transducer
Efficient surface
2.5 cm²
5 cm²
US frequency
1 MHz
3 MHz
1 MHz
3 MHz
Maximum output
7.5 W
2.5 W
15 W
5 W
Effective output
5 W
2.5 W
10 W
5 W
Average power density
3 W/cm²
1 W/cm²
3 W/cm²
1 W/cm²
Modulation
cont. (continuous)
0 ...(0.5...2.0) W/cm² =
cont.
(0.5....2.0)...3.0 W/cm² =
2/5
100 Hz
- 1 / 10
100 Hz
- 1 / 5
100 Hz
- 1 / 3
100 Hz
- 2 / 5
100 Hz
Ambient temperature
+ 10°C ... + 40°C
Dimensions
(W x H x D)
34.5 x 13.3 x 26.7 cm
Weight
3.9 kg